3.14 Data Storage
Store
data from Main Board to HandControl. (Use when Main Board needs to be replaced).
Store
Store
Store
If store failed: - Main Board failure.
- Data storage not possible.
Transfer
data from HandControl to Main Board. Use only on Main Board replacement. Require successfully completed
Data Storage 4.14.
Data
Advanced
Transfer
Transfer
Transfer
Transfer
Data
If transfer failed: - Try one more time
- Make sure Data storage is done
- Make sure all connections are properly
assembled.
- Contact Liko/Hill-Rom
3.15 Data Transfer