1.1 Technical Data
*
Wall charger; full charge is achieved after maximum 6 hours.
**
IRC charger; full charge is achieved after maximum 16 hours.
Maximum load
L: 272 kg (600 Ibs.)
XL: 363 kg (800 Ibs.)
Battery:
Main board memory
backup battery:
25.2V / 4.2 Ah
Button cell battery Lithium 3 V 225 mAh, CR2032
Battery charger:
Built-in battery
Power source:
Wall charger, Art. No. 3305010 - 3305050
*
Input: 100-240 VAC, 50/60 Hz, 0.9 A
Output: 33.5 VDC, 1.36 A
IRC charger, Art. No. 3305510 -3305550
**
Input: 100-240 VAC, 50/60 Hz, 1.2 A
Output: 27.6-29.5 VDC, 1.5 A
Lifting speed:
5-6 cm/sec with a load between 0 - 100 kg
(2 - 2.3 inch./sec. with a load between 0 - 220 Ibs.)
4-6 cm/sec with a load between 100 kg - SWL
(1.6 - 2.3 inch./sec. with a load between 220 Ibs. - SWL)
Lifting interval:
Min. 2300 mm
Electrical data:
25.2 V / 30 A
Fuse:
Type: 0287020 Voltage: 32VDC Breaking capacity: 1000A
Operation Speed: 150ms-5s
Liftmotor Weight:
14 kg (31 Ibs.)
Emergency
lowering device:
Mechanical
Electrical
Emergency raising:
Electrical
Duty cycle:
(Intermittent
power)
10/90 max. 2 min (0 - 272 kg) (0 - 600 Ibs.)
5/95 max. 2 min (.>272 - 363 kg) (.> - 800 Ibs.)
Sound level:
70 dB(A)
Lift motor
Protection class:
IP X4 (humidity resistance)
Hand control
Protection class:
IP X7 (humidity resistance)
Operating forces of
controls:
4.9 N
Surrounding
functional
environment:
T
emp. +5 ° C till +40 ° C, relative humidity 15% to
93% non-condensing.
Barometric pressure range of 70 kPa to 106 kPa
.
Intended for indoor use.
Protection class against electric shock Type BF,
(Wall Charger).
Protection class against electric shock Type B,
(IRC Charger)
.
Top view
Side view
CSP
(Central Suspension Point)
*
Min. distance from ceiling to CSP at max. lifting height.
**
Installation dimensions: the distance between the attachment point for the lift unit on the carriage and the CSP at max. lifting height.
*** Lifting interval: the distance between max. lifting height and min. lifting height measured in CSP.
Measurements in mm.
C
A
B
H**
D*
L***
A
B
C
D*
H**
L***
178
410
250
323
232
2300
Ceiling
The product fulfils the requirements of:
ISO10535:2006, IEC60601-1, IEC60601-1-2, IEC60601-1-6 , IEC60601-1-11, ANSI/AAMI ES60601-1, CAN/CSA C22.2 No 60601-1
WARNING
!
This product shall only be used in the recommended environment.
Measurements in Inch.
A
B
C
D*
H**
L***
7
16.1
9.8
12.7
9.1
90.6