![Hill-Rom 3301030 Service Manual Download Page 10](http://html1.mh-extra.com/html/hill-rom/3301030/3301030_service-manual_2134154010.webp)
Service Manual LikoGuard™
overhead lift
• 3EN190401 Rev. 7
10
www.hillrom.com
3.1 Assembly of HandControl with display, for service
HandControl LG 2-Button with display
Product No. 3306020
kg/lbs
When the HandControl with display is required for service and the HandControl without display is attached to
LikoGuard™ Overhead Lift, this assembly instruction can be used:
Step 1.
-Activate the emergency stop
-Remove side cover (Screwdriver)
Step 2.
-Release the HandControl cables to access the cover.
(White cable)
-Remove the cover by slideing it up.
Step 4.
-Disconnect the HandControl without display.
Step 3. (If Step-up card is installed)
-Remove the board, 2xT15 screws.
Before assembly:
•
ESD protection is required.
Step 5.
-Connect the HandControl with display and start
the service.
Tools required:
T15