42
When a capillary skin puncture is performed, several
defense systems in the body are activated very quickly.
These defense systems cause an increase in the number
of WBCs in the blood closest to the wound, leading to
greater differences in results from several samples taken
from the same finger stick.
A study was performed, on the HemoCue WBC system in
which possible influences of capillary sampling technique
were included, see Figure 3 below.
Figure 3
Capillary Sampling Technique
An additional study has been performed at a clinical
reference laboratory in the U.S. including a large number
of blood samples in order to evaluate the HemoCue WBC
system for expected demographic variations (age, gender,
geographics) in the population, see Figure 4 below.
Figure 4
Demographic Sample Population
H
em
o
C
u
e W
B
C
, x
10
9
/L
Sysmex XS-1000i, x10
9
/L
-------
X = Y
Coefficients
b
0
= 0.066
b
1
= 1.033
r
2
= 0.972
N = 119
H
em
o
C
u
e W
B
C
, x
10
9
/L
Beckman Coulter, x10
9
/L
Coefficients
b
0
= – 0.451
b
1
= 0.970
r
2
= 0.994
N = 178
-------
X = Y
Summary of Contents for WBC
Page 1: ...HemoCue WBC Operating Manual Manuel d utilisation ...
Page 2: ......
Page 4: ......
Page 6: ...HemoCue CleanerWBC 3 2 1 4 5 6 6 Components Éléments ...
Page 8: ...1 2 3 8 Start up Mise en service ...
Page 10: ...4 5 6 7 4 10 ...
Page 12: ...1 2 3 12 Set up Audio signal Configuration Signal sonore ...
Page 14: ...1 2 4 3 14 Measuring Capillary blood Mesure Sang capillaire ...
Page 16: ...5 7 6 8 16 ...
Page 18: ...10 9 11 18 ...
Page 20: ...12a 12b 13 14 15 20 ...
Page 24: ...1 2 3 24 Set up Printer function Configuration Mode impression ...
Page 26: ...1 2 3 4 5 26 Maintenance Daily Maintenance Entretien Entretien quotidien ...
Page 28: ...1 2a 2b 3 4 28 Maintenance Optical Parts Entretien Pièces optiques ...
Page 66: ...66 ...
Page 67: ...67 ...