40
Specific Performance Characteristics
Linearity
The method used in the HemoCue WBC system has
accor ding to the CLSI Document EP6-A been dem-
onstrated to be linear between 0.3–30.0 x10
9
/L
(300–30000/mm
3
, 300–30000/µL), within 0.2 x10
9
/L
difference in 0.3–3.5 x10
9
/L and within 6 % difference
in 3.6–30.0 x10
9
/L.
Within-run and Total Precision
Within-run and Total Precision was determined accor ding
to the CLSI Document EP5-A2.
The results given below
come from four batches of HemoCue WBC Microcu-
vettes, and five HemoCue WBC Analyzers. Commercially
available controls at three different levels were used. The
WBC count was measured in duplicate twice a day, morn-
ing and afternoon, during 20 consecutive days.
Level
N
x10
9
/L
Within-run
Precision
CV %
Total
Precision
CV %
1
400
2.5
4.1
5.4
2
400
7.2
2.9
3.5
3
400
19.0
1.6
1.9
Additionally, a study has been performed to determine
the precision using fresh blood samples. The study was
performed on venous whole blood samples tested in
duplicate.
x10
9
/L
N
CV
duplicates
%
Sysmex XS-1000i
HemoCue WBC
0.3–1.0
12
5.0
10.4
1.1–3.5
52
3.7
4.1
3.6–10.0
209
1.9
3.5
10.1–20.0
88
2.1
2.5
20.1–30.0
35
1.8
2.2
Total
396
2.4
3.2
Summary of Contents for WBC
Page 1: ...HemoCue WBC Operating Manual Manuel d utilisation ...
Page 2: ......
Page 4: ......
Page 6: ...HemoCue CleanerWBC 3 2 1 4 5 6 6 Components Éléments ...
Page 8: ...1 2 3 8 Start up Mise en service ...
Page 10: ...4 5 6 7 4 10 ...
Page 12: ...1 2 3 12 Set up Audio signal Configuration Signal sonore ...
Page 14: ...1 2 4 3 14 Measuring Capillary blood Mesure Sang capillaire ...
Page 16: ...5 7 6 8 16 ...
Page 18: ...10 9 11 18 ...
Page 20: ...12a 12b 13 14 15 20 ...
Page 24: ...1 2 3 24 Set up Printer function Configuration Mode impression ...
Page 26: ...1 2 3 4 5 26 Maintenance Daily Maintenance Entretien Entretien quotidien ...
Page 28: ...1 2a 2b 3 4 28 Maintenance Optical Parts Entretien Pièces optiques ...
Page 66: ...66 ...
Page 67: ...67 ...