30
Symptom
Explanation
Action
Err01
A portion of the image area is unable to be
analyzed.
1. Due to measurement error.
2. Due to abnormal sample.
1. Take a new microcuvette and repeat the
measurement, as described in the Measuring section.
2. NOTE! The blood sample should be verified with a
suitable laboratory method and be questioned as to
the pathological condition of the patient.
Err02
Uneven spatial distribution of detected
cells.
Take a new microcuvette and repeat the measurement,
as described in the Measuring section.
Err03
Image, or part of the image area is detected
as out-of-focus.
Take a new microcuvette and repeat the measurement,
as described in the Measuring section.
Err30
1. Optical parts dirty.
2. Optical parts wet after cleaning.
1. Turn off the analyzer and clean the optical parts as
described in the Maintenance section.
2a. Wait 15 minutes before turning on the analyzer
after cleaning to make sure that the optical parts are
dry.
2b. If the problem continues, the analyzer needs
service. Contact your local distributor.
Err33
Empty microcuvette, not filled with sample.
Take a new microcuvette and repeat the measurement,
as described in the Measuring section and make sure
that the microcuvette is filled with sample.
Err34
Stray light detected.
a. Turn off the analyzer and make sure the analyzer is
not exposed to any bright light sources.
b. If the problem continues, the analyzer needs service.
Contact your local distributor.
Err35
The battery power is too low.
Turn off the analyzer.
a. Replace the batteries, six type AA, as described in
the Start Up section.
b. Use the power adapter as described in the Start Up
section.
Troubleshooting Guide
GB/US
If you are unable to resolve the problem by following this Troubleshooting Guide, please contact your local HemoCue distributor or HemoCue AB. The
analyzer should be cleaned as described in the Maintenance section prior to service or disposal. Consult local environmental authorities for proper disposal.
The analyzer has no serviceable parts. NOTE! Do not open the cover of the analyzer. The warranty is void if the cover of the analyzer has been opened.
Summary of Contents for WBC
Page 1: ...HemoCue WBC Operating Manual Manuel d utilisation ...
Page 2: ......
Page 4: ......
Page 6: ...HemoCue CleanerWBC 3 2 1 4 5 6 6 Components Éléments ...
Page 8: ...1 2 3 8 Start up Mise en service ...
Page 10: ...4 5 6 7 4 10 ...
Page 12: ...1 2 3 12 Set up Audio signal Configuration Signal sonore ...
Page 14: ...1 2 4 3 14 Measuring Capillary blood Mesure Sang capillaire ...
Page 16: ...5 7 6 8 16 ...
Page 18: ...10 9 11 18 ...
Page 20: ...12a 12b 13 14 15 20 ...
Page 24: ...1 2 3 24 Set up Printer function Configuration Mode impression ...
Page 26: ...1 2 3 4 5 26 Maintenance Daily Maintenance Entretien Entretien quotidien ...
Page 28: ...1 2a 2b 3 4 28 Maintenance Optical Parts Entretien Pièces optiques ...
Page 66: ...66 ...
Page 67: ...67 ...