39
Recommended separation distance between Portable and mobile RF
communications equipment and HemoCue
®
Hb 301 Analyzer
The HemoCue systems are intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled.
The customer or the user of HemoCue systems can help prevent
electromagnetic interference by maintaining a minimum distance
between portable and RF communications equipment (transmitters) and
HemoCue systems as recommended below, according to the maximum
output power of the communications equipment.
Rated
maximum
output power
of transmitter
(W)
Separation distance according to frequency of
transmitter (m)
150 kHz
to 80 MHz
d=1.2
√
P
80 MHz to
800 MHz
d=1.2
√
P
800 MHz to
2.5 GHz
d=2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at maximum output power not listed above, the
recommended separation distances (d) in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter,
where (P) is the maximum output power rating of the transmitter in
watts (W) according to transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the
higher frequencyrange applies.
NOTE 2
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Summary of Contents for Hb 301
Page 1: ...HemoCue Hb 301 Operating manual Manuel d utilisation ...
Page 2: ......
Page 4: ...4 ...
Page 6: ...QRG CD 1 2 3 4 5 6 Composants 6 Components ...
Page 8: ...1 2 3 4 5 Mise en service 8 Start up ...
Page 10: ...1 2 3 Configuration Signal sonore 10 Set up Audio signal ...
Page 13: ...13 ...
Page 14: ...1 2 3 4 Mesure Sang capillaire 14 Measuring Capillary blood ...
Page 16: ...5 6 7 8 16 ...
Page 18: ...9 10 11 18 ...
Page 20: ...12a 12b 13 14 15 20 ...
Page 24: ...1 2 3 Configuration Fonction PC Impression 24 Set up PC Printer function ...