28
Symptom
Explanation
Action
The analyzer shows an error code.
May be a temporary fault.
Turn off the analyzer and turn it on again after 30 seconds. Take a new micro-
cuvette and repeat the measurement.
If the problem continues, see specific error code below.
E00
No stable endpoint of the measurement is found within the time
range.
1.
The microcuvette is faulty.
2.
The circuit board is out of order.
1a. Check the expiration date for the microcuvettes.
1b. Take a new microcuvette and repeat the measurement.
2.
The analyzer needs service. Contact your distributor.
E01–E05
1.
Dirty optronic unit or faulty electronics or optronic unit.
1a. Turn off the analyzer and clean the optronic unit as descriped in the
maintance section.
1b. The analyzer needs service. Contact your distributor.
E06
1.
Unstable blank value. The analyzer might be cold.
1.
Turn off the analyzer and allow it to reach room temperature. If the problem
continues, the analyzer needs service. Contact your distributor.
E07
1.
The battery power is too low.
1a. The batteries need to be replaced. Turn off the analyzer and replace the
batteries, 4 type AA.
1b. Use the power adapter.
E08
The absorbance is too high.
1.
Light blocking item in the cuvette holder
1a. Check that the analyzer and microcuvettes are used according to the
HemoCue Hb 301 operating manual and instructions for use.
1b. The analyzer needs service. Contact your distributor.
Troubleshooting Guide
GB/US
If you are unable to resolve the problem by following this Troubleshooting Guide, please contact your local HemoCue distributor or HemoCue AB. Consult local environmental authorities for
proper disposal. The analyzer has no serviceable parts. Do not open the cover of the analyzer.
Note: The warranty is void if the cover of the analyzer has been opened.
Summary of Contents for Hb 301
Page 1: ...HemoCue Hb 301 Operating manual Manuel d utilisation ...
Page 2: ......
Page 4: ...4 ...
Page 6: ...QRG CD 1 2 3 4 5 6 Composants 6 Components ...
Page 8: ...1 2 3 4 5 Mise en service 8 Start up ...
Page 10: ...1 2 3 Configuration Signal sonore 10 Set up Audio signal ...
Page 13: ...13 ...
Page 14: ...1 2 3 4 Mesure Sang capillaire 14 Measuring Capillary blood ...
Page 16: ...5 6 7 8 16 ...
Page 18: ...9 10 11 18 ...
Page 20: ...12a 12b 13 14 15 20 ...
Page 24: ...1 2 3 Configuration Fonction PC Impression 24 Set up PC Printer function ...