29
Symptom
Explanation
Action
E10–E30
1.
Dirty optronic unit or faulty electronics or optronic unit.
1a. Turn off the analyzer and clean the optronic unit as described in the maint-
ance section.
1b. The analyzer needs service. Contact your distributor.
E40
1.
The cuvette holder is not replaced properly after cleaning.
2.
Dirty optronic unit.
3.
The microcuvette is not a HemoCue Hb 301 Microcuvette.
4.
The microcuvette is damaged.
1.
Make sure that the cuvette holder is replaced properly as descriped in
the maintance section.
2.
Turn off the analyzer and clean the optronic unit as described in the
maintance section.
3.
Only use HemoCue Hb 301 Microcuvettes in the HemoCue Hb 301 Analyzer.
4.
Take a new microcuvette and repeat the measurement.
E41
1.
Incorrect adapter.
2.
Bad batteries.
3.
Hardware error.
1.
Use an adapter listed under Adapters in Specifications.
2.
Replace batteries.
3.
The analyzer needs service. Contact your distributor.
E42–49
1.
The optronic unit has been scratched due to incorrect
maintance.
2.
Hardware error.
1.
Clean the optronic unit as described in the maintance section, using the
HemoCue Cleaner. The analyzer needs service. Contact your distributor.
2.
The analyzer needs service. Contact your distributor.
HHH
1.
Measured value exceeds 25.6 g/dL
(256 g/L, 15.9 mmol/L).
No characters on the display.
1.
The analyzer is not receiving power.
2.
If on battery power, the batteries need to be replaced.
3.
The display is out of order.
1a. Check that the power adapter is connected to the analyzer and the AC
power supply.
1b. Check that the cable is not damaged.
2.
Turn off the analyzer and replace the batteries, 4 type AA.
3.
The analyzer needs service. Contact your distributor.
The display contains erroneous
characters.
1.
The display is out of order.
2.
The microprocessor is out of order.
1.
The analyzer needs service. Contact your distributor.
2.
The analyzer needs service. Contact your distributor.
Summary of Contents for Hb 301
Page 1: ...HemoCue Hb 301 Operating manual Manuel d utilisation ...
Page 2: ......
Page 4: ...4 ...
Page 6: ...QRG CD 1 2 3 4 5 6 Composants 6 Components ...
Page 8: ...1 2 3 4 5 Mise en service 8 Start up ...
Page 10: ...1 2 3 Configuration Signal sonore 10 Set up Audio signal ...
Page 13: ...13 ...
Page 14: ...1 2 3 4 Mesure Sang capillaire 14 Measuring Capillary blood ...
Page 16: ...5 6 7 8 16 ...
Page 18: ...9 10 11 18 ...
Page 20: ...12a 12b 13 14 15 20 ...
Page 24: ...1 2 3 Configuration Fonction PC Impression 24 Set up PC Printer function ...