360077-1/O
45
i.Series® - All Models
10.6.2
Install the Unit Cooler Cover
1
Verify unit cooler wiring is connected and routed correctly.
a
Wiring should be routed above copper tube inside the unit cooler.
b
Reconnect wires if they have separated.
2
Attach unit cooler cover.
a
Lift unit cooler cover into place.
b
Front edge of the cover should be behind the unit cooler case.
c
Use the socket wrench to install four screws to secure the unit cooler cover.
3
Insert the drain hose through hole in the refrigerator.
a
Push drain hose upward, toward the unit cooler drain port.
b
In the chamber, push drain hose onto unit cooler drain port.
4
On the back of the chamber, press putty around the drain hose.
5
Reinstall the top drawer, shelf, or basket.
6
Reinstall the battery in the monitoring system backup battery holder. Reconnect the refrigerator to AC
power.
7
Press the
Mute
button to disable the high temperature alarm while refrigerator reaches operating
temperature.
10.7 Supplies
Refrigerant: non-CFC, R-134A
Chart paper: 220366 (52 sheets)
Glycerin solution: 400922-1
Fluorescent lamp: T5, 13 W
Monitoring system batteries: (6) 1.5 V, D-cell non-rechargeable alkaline batteries (or equivalent): 715031
Chart recorder battery (optional): (1) 9 V non-rechargeable alkaline (or equivalent): 120218