360077-1/O
74
Horizon Series™ - Blood Bank Models
16.7
Alarm Settings
The following alarm settings may be changed by the operator. The setpoint for temperature alarms may
be changed (where applicable), as well as the time delay between when the alarm condition commences
and when the visual and audible alarms are initiated.
16.7.1
Alarm Volume
The alarm volume can be changed. The Alarm Volume controls volume for all audible alarms.
►
Default setting is 10
►
Setting can be changed from 1 to 10
►
1 is the quietest setting; 10 is the loudest setting
Change the alarm volume:
1
Press the
MAIN
button.
2
Press the
DOWN
button to highlight Edit Configuration. Press the
SELECT
button.
3
Enter the password when prompted.
4
Press the
DOWN
button to highlight System Options. Press the
SELECT
button.
5
Press the
DOWN
button to highlight Alarm Volume.
6
Press the
INC
or
DEC
buttons to change the setting.
7
Press the
BACK
button to return to the Edit Configuration screen, or press the
HOME
button to exit.
The new settings are saved.
16.7.2
Alarm Pulse
The alarm pattern can be changed. This is useful if several refrigerators with alarms are collocated, and
distinguishing the source of the alarm quickly is desirable.
►
Default setting is Single.
►
Setting can be changed between Single, Double, Triple, and Constant.
Change the alarm pulse:
1
Press the
MAIN
button.
2
Press the
DOWN
button to highlight Edit Configuration. Press the
SELECT
button.
3
Enter the password when prompted.
4
Press the
DOWN
button to highlight System Options. Press the
SELECT
button.
5
Press the
DOWN
button to highlight Alarm Pulse.
6
Press the
INC
or
DEC
buttons to change the setting.
7
Press the
BACK
button to return to the Edit Configuration screen, or press the
HOME
button to exit.
The new settings are saved.
16.7.3
High Chamber Temperature Alarm
The High Alarm setpoint specifies the temperature at which the High Temperature Alarm activates. If the
temperature detected by the chamber probe is greater than or equal to this value, the alarm activates.
►
Default setpoint is 5.5 °C
►
Setpoint can be changed from -40 °C to +40 °C
Change the setpoint:
1
Press the
MAIN
button.
2
Press the
DOWN
button to highlight Edit Configuration. Press the
SELECT
button.
3
Enter the password when prompted.
4
Press the
DOWN
button to highlight Alarm Setpoints. Press the
SELECT
button.
5
Press the
DOWN
button to highlight High Alarm Setpoint.
6
Press the
INC
or
DEC
buttons to change the setting.