360077-1/O
33
i.Series® - All Models
Test the alarm:
1
Ensure the monitoring system backup batteries are installed.
2
Remove one battery from the monitoring system backup battery holder.
a
If the no battery alarm activates, no further action is needed. Reinstall the battery.
b
If the no battery alarm does not activate, contact Helmer Technical Service.
NOTICE
When installing replacement batteries, use only batteries which meet the specifications
outlined in chapter
10.7
(Supplies).
8.10
Additional System Settings
8.10.1
Screen Contrast
The screen contrast can be changed for easier viewing.
NOTE
►
During an AC power failure, the screen backlight is not illuminated to conserve
backup battery power.
►
During an AC power failure, the screen contrast cannot be changed.
Change screen contrast:
1
On the HOME screen, press the third button from the left to make the text appear lighter.
2
On the HOME screen, press the second button from the left to make the text appear darker.
8.10.2
Date and Time
The Date Format setting controls the order in which the month (mm) and day (dd) are displayed.
►
Month is a 2-digit number (01-12)
►
Day is a 2-digit number (01-31)
►
Default date format is mm/dd/yyyy
The Clock Mode setting controls whether the time is displayed in a 12-hour or 24-hour format.
►
When using the 12-hour format, AM or PM must be specified
►
Default setting is 12-hour
Change date and time settings:
1
Press the
MAIN
button.
2
Press the
DOWN
button to highlight Edit Configuration. Press the
SELECT
button.
3
Enter the password when prompted.
4
Press the
DOWN
button to highlight Set Date & Time. Press the
SELECT
button.
5
Press the
UP
or
DOWN
buttons to select the date and time settings to change.
6
Press the
INC
or
DEC
buttons to change the setting.
7
Press the
BACK
button to return to the Edit Configuration screen, or press the
HOME
button to exit.
The new settings are saved.
8.10.3
Display Language
The i.Center monitoring system stores two languages. English is the default language. If a different
language is desired, it must be loaded from the flash memory card. If a flash memory card is not included
with the refrigerator, the languages may have been loaded to the i.Center prior to shipment.
To obtain a flash memory card to load an alternate language, contact Helmer Technical Service.
NOTE
Each time the refrigerator is powered on, the i.Center display language must be selected.
Summary of Contents for (Version A) HB120(Versions A and B)
Page 137: ......