i.Series® - All Models
Section II: i.Series
®
- All Models
NOTE
This section applies to iB, iPR, iLR, and iHB models.
7
Product Configuration
7.1
Install Battery for Backup Power
The monitoring system and chart recorder each have a battery system, enabling a period of continuous
operation if power is lost.
NOTE
The monitoring system will not start on battery power alone. If the refrigerator was
previously not connected to AC power and the batteries are installed, the monitoring
system will not run on battery power.
Battery life varies by manufacturer as well as voltage level remaining. Providing full power is available
and no battery-related alarms are active, backup power for the monitoring system is available for up to
two hours.
NOTICE
When installing replacement batteries, use only batteries which meets the specifications
outlined in chapter
10.7
(Supplies).
The batteries are located on the top of the refrigerator, behind the monitoring system. For 111 models,
a service cover covers the components and an access panel provides access to the monitoring system
backup batteries.
Monitoring system backup batteries.
Five batteries are installed and one battery is included in the accessory package. Install the sixth battery
to provide power to the monitoring system in the event of an AC power failure.