Horizon Series™ - Blood Bank Models
18.5.5
Clean and Refill Probe Bottle
NOTE
A kit that includes a probe bottle and glycerin is available from Helmer.
Obtain:
►
Fresh water-bleach solution (not provided)
►
1:9 ratio of bleach to water
►
Bleach is 5% solution of commercial sodium hypochlorite (NaOCl)
►
Equivalent oxidizing cleaner/disinfectant approved by your organization may be substituted
►
4 oz. (120 mL) of product simulation solution per bottle
►
10:1 ratio of water to glycerin
Clean and refill bottle:
1
Remove probe from bottle.
2
Remove bottle from bracket.
3
Clean bottle with water-bleach solution.
4
Fill bottle with 4 oz. (120 mL) of product simulation solution.
5
Cap bottle tightly to minimize evaporation.
6
Place bottle in bracket.
7
Replace probe, immersing at least 2” (50 mm).
18.6
Unit Cooler Cover Removal and Installation
If unit cooler cover is not removed as detailed in this procedure the drain port may be damaged. Improper
drainage may result in excessive icing and refrigerator’s inability to maintain temperature.
Required tools:
►
5/16” socket wrench
►
Tool to push putty away from the drain hose
A
B
C
Drain line and hose.
Label Description
A
Unit cooler cover
B
Drain port
C
Drain hose