360077-1/O
43
i.Series® - All Models
10.5.3 Interior
Clean painted surfaces with mild detergent. Clean stainless steel surfaces with a general-purpose
laboratory cleaner suitable for stainless steel.
10.5.4
Door Gaskets
Clean with soft cloth and mild soap and water solution.
10.5.5
Clean and Refill Probe Bottle
NOTE
A kit that includes a probe bottle and glycerin is available from Helmer.
Obtain:
►
Fresh water-bleach solution (not provided)
►
1:9 ratio of bleach to water
►
Bleach is 5% solution of commercial sodium hypochlorite (NaOCl)
►
Equivalent oxidizing cleaner/disinfectant approved by your organization may be substituted
►
4 oz. (120 mL) of product simulation solution per bottle
►
10:1 ratio of water to glycerin
Clean and refill bottle:
1
Remove probe from bottle.
2
Remove bottle from bracket.
3
Clean bottle with water-bleach solution.
4
Fill bottle with 4 oz. (120 mL) of product simulation solution.
5
Cap bottle tightly to minimize evaporation.
6
Place bottle in bracket.
7
Replace probe, immersing at least 2” (50 mm).