360077-1/O
23
i.Series® - All Models
8.2
Main Screen
The Main screen displays functional options that allow access to all other screens in the system.
MAIN screen functional options.
Functions available from the Main screen:
Option
Function
Event Log
►
View historical information about alarms and operational events
System Alarm Test and Status
►
Start or stop an automatic test for temperature alarms
►
View the number of days remaining before the paper for the
temperature chart recorder needs to be changed
►
View the current status of the door (OPEN or CLOSED)
►
View the current condenser temperature
Edit Configuration (password
required)
►
Change the language used for text
►
Change date and time information
►
Change temperature units
►
Change the volume and pattern for audible alarms
►
Enable or disable the chart paper timer
►
Enable or disable the temperature graph display
►
Change alarm-related setpoints and timers
►
Calibrate the temperature probe reading
►
Change some settings to the factory default values
►
Change the password, preventing unauthorized changes
View Configuration
►
View the date and time formats
►
View alarm-related setpoints and timers
►
View the volume and pattern for audible alarms
►
View the setting for the chart paper timer
►
View the setting for the temperature graph display
►
View the settings for temperature and time alarms
Product/Company Information
►
View the software versions for control and display components of
the monitoring system
►
View information to contact Helmer
i.Help
►
Access the on-board help system
Summary of Contents for (Version A) HB120(Versions A and B)