background image

USER MANUAL

EN

version:002-6-2020

Warning: 
Before using this product, 
please read this manual carefully 
and keep it for future reference. 

The design and specifications 

are subject to change without 
prior notice for product 
improvement. Consult with your 
dealer or manufacturer for 
details.

PHARMACY 
REFRIGERATOR

Product implementation stardards:
EN 61010-1 

2010/A1:2019

EN 61010-2-011:2017
EN 61326-1 :2013

Model:VS50

Summary of Contents for VS50

Page 1: ...keep it for future reference The design and speci cations are subject to change without prior notice for product improvement Consult with your dealer or manufacturer for details PHARMACY REFRIGERATOR...

Page 2: ...ries 6 3 PRODUCT OVERVIEW 7 4 TECHNICAL SPEC 7 5 OPERATION 8 5 1 Opera on and display panel 8 5 2 Key opera on 9 5 3 Noise Inside the Unit 13 5 4 ALARM 13 6 CLEANING 14 6 1 Cleaning the Interior and t...

Page 3: ...building in clear of obstruction WARNING Do not use mechanical devices or other means to accelerate the defrosting process other than those recommended by the manufacturer WARNING Do not damage the r...

Page 4: ...r more must be used and copper core conductors in the wall connected to the power socket must also have a cross sectional area of 4mm2 or more This series of products are strictly prohibited to put in...

Page 5: ...socket are plugged firmly and reliably otherwise it may lead to fires When unplugging from the power socket hold the power plug firmly and remove it Do not pull the wires of the power plug directly If...

Page 6: ...he insulation and cause leakage or electric shock Do not insert metal objects such as nails or wires into any openings and gaps of the Pharmacy Refrigerator or into any vent used for internal air circ...

Page 7: ...at both sides 15 cm at the rear and 20 cm at the top of the unit This will allow cold air to circulate around the Pharmacy Refrigerator and improve the efficiency of the cooling process The Pharmacy R...

Page 8: ...to the operation instructions below for details 4 Set parameter of cabinet temperature high low temperature alarm deviation etc according to actual need Default temperature setting is 5 C high temper...

Page 9: ...Model number VS50 External dimensions W D H mm 485 450 550 Effective volume 42L Door body Transparent glass door Foaming agent Cyclopentane Cooling mode Air cooling circulation Shell plate Inner Plat...

Page 10: ...on Button Locked off Button unlocked Power failure on Power off off Power on Printer disabled on Printer working off Printer not working Door heat disabled on Door heat working off Door heat not worki...

Page 11: ...locked if there is no operation for 60s or Up key and Down key are pressed for 3s at the same time b When the key is locked only ambient temperature can be checked and Mute key and Light switch key c...

Page 12: ...Lower deviation on low temperature alarm 0 0 10 0 0 cancel the alarm 3 n Check the current date year y Check the current date month r Check the current date day S Check the current date hour F Check t...

Page 13: ...o 1 Every time the door is opened and closed it will be heated for T9 minutes If the door is opened and closed again during the heating period the heating time will be reset Door heating mode P1 is se...

Page 14: ...eps once the digital tube displays the upload data code on in case of large amount of data and the USB flash disk generates the PDF files on data of current month and last month After the completion o...

Page 15: ...bient temperature alarm value tH1 sensor is normal and not shielded then the buzzer will alarm and the digital tube will flash the high ambient alarm code H2 alternately at the interval of 3 seconds e...

Page 16: ...fied technician for assistance 7 MAINTENANCE 7 1 Care When Handling Moving Your Unit Hold the unit around its sides or base when moving it Under no circumstances should it be lifted by holding the edg...

Page 17: ...t heat during operation specially in summer this is caused by the radiation of the condenser and it is a normal phenomenon Condensation condensation phenomenon will be detected on the exterior surface...

Page 18: ...16 REQUESTING ASSISTANCE OR SERVICE If you have any doubt or extra assistance please call Customer Service Vacc Safe Telephone 1300 459 140...