ERS COMPACT
2.3.3
Installing the device
•
Select the location for the device so that the connection to the load resp. source is as short
as possible.
•
Leave sufficient space, at least 30 cm (1 ft), behind the equipment for ventilation (only required
for the standard air-cooled version)
•
The device must not be operated without a proper touch protection for the AC connection,
which is either only accomplished by installation of the device in a 19” rack/cabinet with lockable
doors or by applying further measures (additional cover etc.)
A device in a 19” chassis will usually be mounted on suitable rails and installed in 19” racks or cabinets. The depth
of the device and its weight must be taken into account. The handles on the front are for sliding in and out of the
cabinet. Slots on the front plate are provided for fixing the device (fixing screws not included).
The unacceptable positions, as shown below, are also valid for the vertical mount of the device onto a wall or inside
a cabinet. The required air flow would be insufficient. This wouldn’t actually apply to water cooled models, but the
construction of the entire device isn’t made for operation in this position.
Acceptable and unacceptable installation positions (air or water cooled, air cooled shown):
Standing surface