background image

 

 

TROUBLESHOOTING 

Refer to the following instructions to check and correct any failure before contacting service personnel. 

ERROR LIST 

CALIBRATION PROCEDURE  

This scale has been calibrated at the factory and does not require re-calibration prior to use. 

The following procedure should be performed only to correct any error in calibration. For maximum accuracy, calibration should be done in kilograms 
using certified weights. If you only have certified weights in pounds, convert them to kilograms (i.e. 1 lb = 0.454 kg). The recommended calibration 
weight is 200 kg, however calibration can be performed using any weight no less than 20 kg and no more than 200 kg. 

1.  Turn the scale on.  
2. 

At the same time press the BACKLIGHT and TARE / ZERO buttons. “t” and the a number representing the current automatic shut of

f value will  

appear on the display. 

3. 

Press the HOLD / RELEASE button. “CAL” will appear on the display.

 

4. 

Press the HOLD / RELEASE button again. The display will show “00000.0” and “▼” above HOLD.

 

5.  Using the TARE / ZERO and KG / LB buttons to move the arrow right or left respectively and the BACKLIGHT button to change each digit, set the 

desired calibration weight.  

6.  Load the calibration weight on the scale and press the HOLD / RELEASE button. The calibration weight loaded on the scale and an arrow above 

STABLE will appear on the display. 

7.  Remove the weight load from the scale and the scale returns to its normal weighing mode.  

SYMPTON 

POSSIBLE CAUSE 

CORRECTIVE ACTION 

Scale does not turn on  

1.  Dead Battery  
2.  Faulty electrical outlet  
3.  Bad Power Supply 

1.  Attach adapter to recharge battery.  
2.  Use a different outlet.  
3.  Replace adapter  

Questionable weight or the scale 
does not zero  

1.  External object interfering with the scale.  
2. 

The display did not show „0.0‟ before 

weighing  

3.  Scale is not on a level floor  

 

4.  Scale is out of calibration  

1.  Remove interfering object from the scale.  
2.  Ask the patient to get off the scale and begin weighing 
 

process again.  

3.  Place the scale on a level floor and begin weighing process 
 

again.  

4.  Check weight with know weight value.  

The display shows “ 

- 0F- 

“ 

 

The load on the scale exceeds the maximum 
capacity (440 lbs / 200 kg)  

Remove the excess weight and use the scale according to its 
limits.  

ERROR NO. 

DEFINITION 

CHECK / REPLACE  

Err 1  

A / D value is lower when CAL  

Scale is below minimum calibration weight. Calibration weight 
must be more than 20kg. The recommended calibration weight 
is 200kg (440 lbs). 

--OF--  

The load on the scale exceeds the capacity (440 lb / 200kg).  Remove excess weight and use the scale according to its 

limits. 

0-H  

High saturation state (High A / D)  

The load cell is not connected properly. Check cables and 
mechanical connection. If problem persist, replace the load 
cell. 

0-L  

Low saturation state (Low A / D)  

The load cell is not connected properly. Check cables and 
mechanical connection. If problem persist, replace the load 
cell. 

P-L  

Battery is low.  

Recharge or replace battery.  

Summary of Contents for 594kl

Page 1: ...O ELECTRONIC CHAIR SCALE MODEL 594KL FAUTEUIL DE PESÉE ÉLECTRONIQUE MODÈLE 594KL BALANZA ELECTRONICA CON SILLA MODELO 594KL Please register your product at www homscales com Veuillez enregistrer votre produit sur www homscales com Registre su balanza en nuestro sitio Web www homscales com Rev 20100722 ...

Page 2: ...use Do not use in the presence of flammable materials Operating at other voltages and frequencies than specified could damage the equipment Do not leave patient unattended in the chair Do not use chair scale in place of a wheelchair to transport patients Do not bounce scale down stairs as doing so may damage internal parts SCALE SPECIFICATIONS GENERAL Health o meter s Digital Chair Scale model 594...

Page 3: ...t by the factory is 2 minutes 1 Press ON OFF and wait for the display to show 0 0 2 At the same time press both the BACKLIGHT and TARE ZERO buttons The display will show t along with a number representing the current shut off setting 3 Use the BACKLIGHT button to change the shut off setting The shut off values are set in 2 minute increments t 0 disabled t 1 2 minutes t 2 4 minutes t 9 18 minutes 4...

Page 4: ...load from the scale and the scale returns to its normal weighing mode SYMPTON POSSIBLE CAUSE CORRECTIVE ACTION Scale does not turn on 1 Dead Battery 2 Faulty electrical outlet 3 Bad Power Supply 1 Attach adapter to recharge battery 2 Use a different outlet 3 Replace adapter Questionable weight or the scale does not zero 1 External object interfering with the scale 2 The display did not show 0 0 be...

Page 5: ...r or alternations Further the warranty does not cover Acts of God such as fire floods hurricanes and tornadoes This warranty gives you specifi c legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To get warranty service make sure you keep your sales receipt or document showing proof of purchase Call ...

Page 6: ...un fauteuil roulant pour déplacer les patients Ne descendez pas la balance dans un escalier car cela pourrait endommager les pièces internes SPÉCIFICATIONS DE LA BALANCE GÉNÉRALITÉS Le fauteuil de pesée numérique de Health o meter modèle 594KL utilise une technologie à microprocesseur sophistiquée Cet instrument de précision est conçu pour offrir des mesures de poids précises fiables et reproducti...

Page 7: ... écran affiche 0 0 2 Au même moment appuyez sur les boutons BACKLIGHT et TARE ZERO L écran affichera t en plus d un chiffre représentant le réglage d arrêt automatique actuel 3 Utilisez le bouton BACKLIGHT pour modifier le réglage d arrêt automatique Les valeurs d arrêt sont offertes en incréments de deux minutes t 0 désactivée t 1 2 minutes t 2 4 minutes t 9 18 minutes 4 Appuyez sur le bouton HOL...

Page 8: ...ÔME CAUSE POSSIBLE ACTION CORRECTRICE La balance ne se met pas en marche 1 Pile déchargée 2 Prise électrique défectueuse 3 Mauvaise alimentation électrique 1 Branchez l adaptateur pour recharger la pile 2 Utilisez une autre prise 3 Remplacez l adaptateur Poids suspect ou la balance ne se réinitialise pas à zéro 1 Un objet externe fait obstacle à la balance 2 L écran n a pas indiqué 0 0 avant le pe...

Page 9: ... abus y compris l altération les dommages subis en transit ou les réparations ou les modifications non autorisées De plus la garantie ne couvre pas les catastrophes naturelles comme les incendies les inondations les ouragans et les tornades Cette garantie vous confère des droits légaux précis et vous pouvez aussi avoir d autres droits qui varient d un pays à l autre d un état à un autre d une prov...

Page 10: ...r pacientes No haga rebotar la balanza escaleras abajo dado que de hacerlo podría dañar las partes internas ESPECIFICACIONES DE LA BALANZA INFORMACIÓN GENERAL La balanza digital de silla Health o meter modelo 594KL utiliza una compleja tecnología de microprocesadores Este instrumento de pre cisión está diseñado para brindar mediciones del peso exactas confiables y repetibles y posee característica...

Page 11: ... ON OFF y espere a que se muestre 0 0 en la pantalla de lectura 2 Presione simultáneamente los botones BACKLIGHT y TARE ZERO En la pantalla se mostrará t junto a un número que corresponde al valor vigente del apagado 3 Use el botón BACKLIGHT para cambiar la configuración del apagado Los valores de apagado se establecen por múltiplos de 2 minutos t 0 inhabilitado t 1 2 minutos t 2 4 minutos t 9 18 ...

Page 12: ...IÓN CORRECTIVA La balanza no se enciende 1 Batería agotada 2 Tomacorriente defectuoso 3 Fuente de alimentación defectuosa 1 Conecte el adaptador para recargar la batería 2 Use otro tomacorriente 3 Cambie el adaptador El pesaje es dudoso o la balanza no está en cero 1 Un objeto externo interfiere con la balanza 2 En la pantalla no se mostraba 0 0 antes de pesar 3 La balanza no está sobre una superf...

Page 13: ...trato incluida la manipulación indebida del producto los daños sufridos durante el transporte y las reparaciones o modificaciones no autorizadas Además esta garantía tampoco cubre los daños causados por desastres naturales como incendios inundaciones huracanes y tornados La presente garantía le otorga derechos legales específicos y es posible que usted tenga también otros derechos diferentes según...

Page 14: ...E PEDIATRIC NEONATAL PRIMARY CARE PROFESSIONAL HOME HEALTH VETERINARY Please call us with questions or visit our web site to learn more about our full lines of professional scales Pelstar LLC 11800 South Austin Avenue B Alsip IL 60803 USA Www homscales com 1 800 638 3722 or 1 708 377 0600 ...