Error
Possible causes
Remedy
Air is released too fast or too
slowly during the measuring
procedure.
The cuff could have become
detached or loosened.
Movement during the
measuring procedure could also
be the reason.
Error during air release or
unusual reduction in pressure.
Check whether the cuff has
been positioned correctly.
Do not move during the
measurement.
Air tube has not been inserted
into the device correctly or is
kinked.
Check that the connector has
been fitted correctly and ensure
that there are no kinks in the
air tube.
The inflation pressure is higher
than 300 mmHg.
The pressure is released
automatically.
Repeat the measurement after
at least 1 minute of rest.
If the battery symbol is
flashing, the batteries are
almost empty. Only a few more
measurements can be taken.
Have new batteries of the same
type ready (AA/LR06).
If the battery symbol is
permanently displayed, the
batteries are empty and must
be replaced.
Insert new batteries of the
same type (AA/LR06). Then
check the date/time and, if
necessary, reset them (see
Chapter 4).
Implausible
measured values
Implausible measured
values often occur due to
inappropriate handling of the
device or if there are mistakes
in the measuring procedure.
Please note the 10 golden
rules for blood pressure
measurement (see Chapter 5.1)
as well as all the important
instructions in Chapter 2.
Then repeat the measurement.
If implausible values still occur:
please contact your doctor!
IFU_Veroval_dc_SK6_0307662_2020-02-10.indd 32
27.02.20 13:09
Summary of Contents for DC3-18
Page 2: ...2 IFU_Veroval_dc_SK6_0307662_2020 02 10 indd 2 27 02 20 13 08 ...
Page 38: ...Notes 38 EN IFU_Veroval_dc_SK6_0307662_2020 02 10 indd 38 27 02 20 13 09 ...
Page 39: ...Notes 39 EN IFU_Veroval_dc_SK6_0307662_2020 02 10 indd 39 27 02 20 13 09 ...
Page 40: ...40 Türkçe TR 1 IFU_Veroval_dc_SK6_0307662_2020 02 10 indd 40 27 02 20 13 09 ...
Page 78: ...78 Notlar TR IFU_Veroval_dc_SK6_0307662_2020 02 10 indd 78 27 02 20 13 09 ...
Page 79: ...79 Notlar TR 1 IFU_Veroval_dc_SK6_0307662_2020 02 10 indd 79 27 02 20 13 09 ...
Page 80: ...80 ﻋﺮﺑـﻲ AE 1 IFU_Veroval_dc_SK6_0307662_2020 02 10 indd 80 27 02 20 13 09 ...
Page 114: ...114 ﻣﻼﺣﻈﺎت AE IFU_Veroval_dc_SK6_0307662_2020 02 10 indd 114 27 02 20 13 09 ...
Page 115: ...115 ﻣﻼﺣﻈﺎت AE 1 IFU_Veroval_dc_SK6_0307662_2020 02 10 indd 115 27 02 20 13 09 ...
Page 116: ...116 繁體中文 HK 1 IFU_Veroval_dc_SK6_0307662_2020 02 10 indd 116 27 02 20 13 09 ...
Page 148: ...148 IFU_Veroval_dc_SK6_0307662_2020 02 10 indd 148 27 02 20 13 09 ...
Page 150: ...150 IFU_Veroval_dc_SK6_0307662_2020 02 10 indd 150 27 02 20 13 09 ...
Page 151: ...151 IFU_Veroval_dc_SK6_0307662_2020 02 10 indd 151 27 02 20 13 09 ...