39
English
EN
Table 4
For NON-LIFE-SUPPORT MEDICAL-ELECTRICAL (ME) DEVICES
and ME SYSTEMS
Recommended separation distances between portable and mobile RF
communications equipment and the Veroval
®
duo control blood pressure
monitor
The Veroval
®
duo control blood pressure monitor is intended for use in one of the
electromagnetic environments specified below in which radiated RF disturbances are
controlled. The customer or user of the Veroval
®
duo control blood pressure monitor can
help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the Veroval
®
duo
control blood pressure monitor, as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output
power of transmitter W
Separation distance (m) according to transmitter frequency
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.2
1.2
2.3
10
3.7
3.7
7.4
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be determined using the equation in the
relevant column, whereby P is the maximum nominal power of the transmitter in watts
(W) according to the transmitter manufacturer’s specifications.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Subject to change without prior notice
IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918.indd 39
17.10.18 18:04
Summary of Contents for 9255020
Page 1: ...IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 1 17 10 18 18 03 ...
Page 2: ...2 English EN IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 2 17 10 18 18 03 ...
Page 40: ...40 Türkçe TR IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 40 17 10 18 18 04 ...
Page 79: ...79 Notlar TR IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 79 17 10 18 18 04 ...
Page 81: ...81 AE ﻋﺮﺑـﻲ IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 81 17 10 18 18 04 ...
Page 118: ...118 ﻣﻼﺣﻈﺎت AE IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 118 17 10 18 18 04 ...
Page 120: ...120 繁體中文 HK IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 120 17 10 18 18 04 ...
Page 158: ...158 IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 158 17 10 18 18 04 ...