HI 1746-WS WEIGH SCALE MODULE
5-12
FIG. 5-6 C2 LADDER LOGIC EXAMPLE/REMOTE MODE OF
OPERATION
Hard Calibration
Hard Calibration is the traditional method of calibra-
tion that uses test weights. Hardy recommends that
the test weights total 80 to 100% of the scale capacity.
Hard Calibration
Ladder Logic
Example
Step 1.
Check to be sure that the parameters have
been setup for your weighing process. (See
Chapter 4, Setup)
Step 2.
We have provided a Ladder Logic example
explaining how to set the weigh process
parameters. The Ladder Logic example is
meant to provide a ladder logic model
only. Your application may vary and the
example may or may not meet your
requirements.
Step 3.
The Hard Calibration Ladder Logic Exam-
ple is located at the Hardy Web Site. If you
have access to the Internet:
a.
Type the following URL: www.hardyso-
lutions.com
b.
Click on the Support button.
c.
Click on Sample Programs.
Summary of Contents for C Series
Page 8: ...HI 1746 WS WEIGHT SCALE MODULE vi...
Page 10: ...HI 1746 WS WEIGH SCALE MODULE II...
Page 46: ...HI 1746 WS WEIGH SCALE MODULE 4 16 I O Setup Ladder Logic for Remote Mode Operation...
Page 47: ...Chapter 4 Setup 4 17...
Page 54: ...HI 1746 WS WEIGH SCALE MODULE 4 24 SETPARAMS Remote Mode of Operation...
Page 55: ...Chapter 4 Setup 4 25...
Page 56: ...HI 1746 WS WEIGH SCALE MODULE 4 26...
Page 60: ...HI 1746 WS WEIGH SCALE MODULE 4 30...
Page 69: ...Chapter 5 Calibration 5 9...
Page 70: ...HI 1746 WS WEIGH SCALE MODULE 5 10 FIG 5 5 C2 LADDER LOGIC EXAMPLE LOCAL MODE OF OPERATION...
Page 74: ...HI 1746 WS WEIGH SCALE MODULE 5 14...
Page 84: ...HI 1746 WS WEIGH SCALE MODULE 6 10...
Page 92: ...HI 1756 WS MANUAL...