background image

55

54

tion will be given by the interface:

LED

Status

Meaning

SMS enable (green)

blinking

Cellular phone

SMS disable (red)

blinking

error

When this error occurs, the instrument will try repeatedly to
initialize the cellular phone engine and the error will be
disactivated only after a successful initialization.

Note

If SIM card expired or maximum number of remaining mes-
sages reached zero, the only way to disactivate cellular error
is to recharge or substitute the SIM card.

While cellular error is active no remote configuration and PC
communication session can take place until error will be
disactivated. Also SMS service is not available.

BATTERY TEST FAILURE

HI504901

 can test the internal backup battery in order to

check if it is good and correctly connected on the main board.
Interface performs the battery test at device start-up and once a
day at 15:00 (3:00 pm).
If battery test is not successful, then an alarm SMS is submitted
by 

HI504901

 supervisor to the programmed number(s) as

described on the “SMS delivery” section. Also repeated SMSs
are sent if no user confirmation is received.

Note

Every day, after this test procedure, related alarm SMS for bat-
tery test failure will be submitted by the interface even if an user
confirmation was received the day before (if there is still prob-
lem with internal battery). The only way to prevent alarm SMS
sending is to solve the problem related to the internal battery.
While battery problem persists green “SMS enable” and red
“SMS disable” LEDs are in the following conditions:

LED

Status

Meaning

SMS enable (green)

blinking

Battery test failed and

SMS disable (red)

ON

SMS sending enabled

SMS enable (green)

ON

Battery test failed and

SMS disable (red)

blinking

SMS sending disabled

Note

Before removing battery or checking its capability, make sure
that the 

HI504901

 is turned off.

Note

Even if a battery problem persists, all 

HI504901

 functions

are still available.

FAULT CONDITIONS & SELFTEST PROCEDURES

HI504901

 is provided with some selftest procedures to pre-

vent interface problems. The below fault conditions may be
detected by the software:
• EEPROM data error
• Cellular phone error
• Battery test failure
If a problem is found, 

HI504901 

gives to the user a “vi-

sual” indication through green and red LEDs for SMS sending
enable/disable (also for battery test failure an alarm SMS is
submitted to the programmed number(s)).

EEPROM DATA ERROR

EEPROM data error can be detected through automatic
EEPROM test procedure. 

HI504901

 performs this test at start-

up and once every hour.
When an EEPROM data error is detected (i.e. if a checksum
control fails), 

HI504901

 tries to repair the corrupted data. If

it does not succeed, a fault alarm is generated for EEPROM
corruption error. To close this error a new local interface con-
figuration is required.
As soon as an EEPROM error is detected, the following “vi-
sual” LED indication is provided by the interface:

LED

Status

Meaning

SMS enable (green)

ON

EEPROM data

SMS disable (red)

ON

corrupted

Note

While EEPROM data error is active no polling operation is
performed. Remote PC communication and remote interface
configuration can not take place. Only a local PC configu-
ration is allowed in order to let the user to verify EEPROM
stored data and to correct the corrupted ones.

CELLULAR PHONE ERROR

During the normal functioning of the interface if a problem
occurs in communication with the cellular engine, a cellular
error will be switched on. 

HI504901

 performs periodically (every

10 minutes) a cellular test to be sure that the cellular phone
engine is ready and network registered. Moreover, a cellular
error can be generated if the SIM card expiration date is reached
or when the number of remaining messages reached zero.
If cellular error is active, the following “visual” LED indica-

Summary of Contents for HI 504901

Page 1: ...HI 504901 GSM Supervisor Instruction Manual...

Page 2: ...tput DIG OUT 13 RS485 Serial Port 13 Power Supply Input 15 Battery 15 SIM Card Insertion 16 START UP 18 IDLE MODE 19 INTERFACE CONFIGURATION 22 Application Software Installation 22 Starting Configurat...

Page 3: ...failure If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization...

Page 4: ...battery which provides power supply in case of mains power failure PRELIMINARY EXAMINATION Remove the instrument from the packing material and exam ine it carefully to make sure that no damage has occ...

Page 5: ...sealed lead acid 12V 0 8Ah 2 On board battery connector 3 GSM module 4 LED indicators see figure on next page 5 RS232 auxiliary communication port for local PC communication only 6 RS232 channel 1 se...

Page 6: ...ity of electronic devices Unplug and turn off the unit in hospitals or near medical devices like pacemakers or hearing aids The module may interfere with the operation of these devices Unplug and turn...

Page 7: ...is standard suitable for data transmission in noisy environ ments INSTALLATION RS232 AUXILIARY PORT HI504901 is provided with an auxiliary RS232 port D Sub 9 pole female plug for connection to a local...

Page 8: ...p to 1 2 Km typical with 24 AWG cable Loads up to 32 typical Internal termination none Connections The connections for the provided 2 pin RS485 terminal are as shown on the below diagram The instrumen...

Page 9: ...10 hours before turning ON the power switch In this manner the new battery can be charged SIM CARD INSERTION Before using the HI504901 insert a SIM card Press the SIM card holder ejector on the cellul...

Page 10: ...ignal2 green ON START UP Turn ON the power switch 14 on page 8 the red SMS disable LED lights up and will stay in this state until the start up procedure ended and the idle mode is en tered During sta...

Page 11: ...output contact The digital output contact status can be changed by the user only by sending the proper SMS to the interface The below table shows the digital output LED indication depending on the di...

Page 12: ...n through serial cable see Local Configuration section To proceed with the remote configuration select the Remote connection type on HI504901SW by checking the proper box On HI 504901 phone number fie...

Page 13: ...each update with a wrong PIN value an attempt will be lost The SMS s stored in the SIM card will be erased Note If an error occurs on cellular phone before pressing the Up date All button the cellula...

Page 14: ...card is used If the feature is enabled it is necessary to insert a value to proceed with the configuration On Expiration Date field it is possible to set the SIM card expiration date dd mm yy this is...

Page 15: ...ed and this operation lasts about 30 seconds the cel lular phone will be turned off and on again to verify the PIN code correctness 15 days 7 days and 1 day before the expiration date warn ing SMS mes...

Page 16: ...ge Each instrument must have a different value for this field Note If the SIM card PIN number is longer than 4 digits it has to be changed before using it in the HI504901 interface To do this insert t...

Page 17: ...axi mum length of this description is 30 characters Digital output section It is possible to enable or disable the control of the digital output status connected to the HI504901 by checking the proper...

Page 18: ...for instruments connected to the RS232 ports will be INSTRUMENTNAME RS232 NUMBEROFTHERS232 CHANNEL Alarms ALARM SPECIFIC TO THE INSTRU MENT Rem msg NUMBER OF REMAINING MESSAGES The format of alarm me...

Page 19: ...turned off or only active alarms when all active alarms can not be sent within one SMS the maximum number of characters for an SMS is 160 more than one message will be sent each one having as header t...

Page 20: ...to send while SMS de lay item sets the delay time between two subsequent alarm messages INFORMATION SMSs In addition to the alarm messages already described the cellu larphoneinterfacecansendinformat...

Page 21: ...01 or HI8002 controller a third SMS will be sent with the label INFO3 3 This message will contain the got from the controlled instruments and could be requested for example after an alarm message rega...

Page 22: ...ne number s This particular situation is managed as an error occurrence and a confir mation is waited When this happens no more messages will be sent by the interface until the error is disactivated l...

Page 23: ...or indication is given by interface green and red LEDs disable enable blink together To disactivate this error and restore the SMS service it is necessary to re charge the SIM card credit Note Every t...

Page 24: ...messages are temporarily sus pended PC COMMUNICATION HI504901 interface can act as a gateway for a connection through a remote computer With this feature a seamless data connection will be established...

Page 25: ...nd PWD interface password CR and the interface will answer ACK char 0x06 if the password is correct or CAN char 0x18 if password is not correct or NAK char 0x15 if command syntax is wrong and hangs up...

Page 26: ...R has to be sent to cellular phone interface For example if target device is attached on RS232 1 port and its communication rate is 2400 bps then the command C0B CR must be issued The cellular phone i...

Page 27: ...rface configuration and PC communications both lo cal remote are not allowed until mains supply comes back When the mains power supply comes back the cellular phone turns on and cellular initializatio...

Page 28: ...capability make sure that the HI504901 is turned off Note Even if a battery problem persists all HI504901 functions are still available FAULT CONDITIONS SELFTEST PROCEDURES HI504901 is provided with s...

Page 29: ...LED blinking fast cellular engine registered to the net work and a call is in progress LED INDICATORS Find here below a brief explanation of all possible LEDs indi cations of HI504901 interface RS232...

Page 30: ...Only two LEDs green and red LEDs for SMS sending enable disable are available to give information to the user about active errors and no more than one can be shown at the same time For this reason ev...

Page 31: ...chargeorsubstitutetheSIMcard Change the SIM card with one enabled for voice calls Note Some SIM cards can be used both for data and voice calls Please check the networkoperatorforfurtherdetails During...

Page 32: ...with remote device RunHI504901SWapplication software and change the options for SMS sending logic on digitalinput window EnableSMSsendingthroughexternalswitch Can not control digital output contact t...

Page 33: ...d electrical shock do not use this instrument when voltage at the measurement surface exceed 24VAC or 60VDC To avoid damage or burns do not perform any measurement in microwave ovens External cables t...

Page 34: ...67 66 USER NOTES...

Page 35: ...584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 South Africa Tel 011 615 607...

Reviews: