background image

35

34

SMS DELIVERY

HI504901

 is able to send SMSs to one (or two) cellular

phone number(s). 

HI504901

 can send two types of mes-

sage: alarm SMSs (upon an alarm condition at least in one
of the instruments connected to the interface) and informa-
tion SMS (upon specific request from the user).

Note

To use the SMS feature, a SIM card able to make a voice call
must be used.

ALARM SMSs

Alarm messages are submitted by 

HI504901

 to the cellular

phone number(s) when an error occurred on one of the instru-
ments connected to the interface. A message about an alarm
condition in one instrument will be sent by the interface only if
the alarm is active and the specific alarm notification was en-
abled during last configuration, by checking the proper box on
the “message option” window of the selected device (see “Inter-
face configuration” section). If an alarm is active in one
instrument connected to the interface but its notification was not
enabled during last configuration, then no SMS will be sent
(however 

HI504901

 gives indication to the user about the

error condition by turning on the red LED for the RS232 chan-
nel or RS485 port or digital input port where the instrument is
connected).

Note

An alarm SMS can be sent only if SMS sending is user en-
abled through digital input terminal (#8 on page 8), other-
wise no SMS will be issued.
The format of alarm messages for instruments connected to the
RS232 ports will be:
“<INSTRUMENT NAME>; RS232-<NUMBER OF THE RS232
CHANNEL>; Alarms: <ALARM SPECIFIC TO THE INSTRU-
MENT>; Rem msg: <NUMBER OF REMAINING
MESSAGES>”
The format of alarm messages for instruments connected to the
RS485 port will be:
“<INSTRUMENT NAME>; RS485-<RS485 DEVICE AD-
DRESS>; Alarms: <ALARM SPECIFIC TO THE INSTRUMENT>;
Rem msg: <NUMBER OF REMAINING MESSAGES>”
During alarm condition, the field <ALARM SPECIFIC TO THE
INSTRUMENT> on the alarm message will contain the list of

Item

Default value

Telephone #1

empty field

Telephone #2

empty field

Date

from internal RTC

Time

from internal RTC

Remaining SMSs

disabled

Expiration date

disabled

Repeated SMSs

2

SMS delay

10 minutes

Query delay

1 minute

Password

disabled

PIN

empty field

RS232 options

no instrument configured

RS485 options

no instrument configured

Digital input

disabled

Digital output

disabled

open the digital output contact. If the digital output control is
enabled, when an information will be requested to 

HI504901

,

an SMS with the digital output status will be sent.

Note

By disabling this option, it will not be possible to set the digi-
tal output status and no information SMS about the digital
output will be sent.

Note

If the digital output management is disabled, the digital output
will be automatically opened, even if it was previously closed.

Note

If the password is not set, only “CLOSEP” or “OPENP” SMS
must be sent.

 

Default values

The following table lists the default value for previous items,
i.e. the factory values copy from ROM if 

HI504901

 was

never configured.

Summary of Contents for HI 504901

Page 1: ...HI 504901 GSM Supervisor Instruction Manual...

Page 2: ...tput DIG OUT 13 RS485 Serial Port 13 Power Supply Input 15 Battery 15 SIM Card Insertion 16 START UP 18 IDLE MODE 19 INTERFACE CONFIGURATION 22 Application Software Installation 22 Starting Configurat...

Page 3: ...failure If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization...

Page 4: ...battery which provides power supply in case of mains power failure PRELIMINARY EXAMINATION Remove the instrument from the packing material and exam ine it carefully to make sure that no damage has occ...

Page 5: ...sealed lead acid 12V 0 8Ah 2 On board battery connector 3 GSM module 4 LED indicators see figure on next page 5 RS232 auxiliary communication port for local PC communication only 6 RS232 channel 1 se...

Page 6: ...ity of electronic devices Unplug and turn off the unit in hospitals or near medical devices like pacemakers or hearing aids The module may interfere with the operation of these devices Unplug and turn...

Page 7: ...is standard suitable for data transmission in noisy environ ments INSTALLATION RS232 AUXILIARY PORT HI504901 is provided with an auxiliary RS232 port D Sub 9 pole female plug for connection to a local...

Page 8: ...p to 1 2 Km typical with 24 AWG cable Loads up to 32 typical Internal termination none Connections The connections for the provided 2 pin RS485 terminal are as shown on the below diagram The instrumen...

Page 9: ...10 hours before turning ON the power switch In this manner the new battery can be charged SIM CARD INSERTION Before using the HI504901 insert a SIM card Press the SIM card holder ejector on the cellul...

Page 10: ...ignal2 green ON START UP Turn ON the power switch 14 on page 8 the red SMS disable LED lights up and will stay in this state until the start up procedure ended and the idle mode is en tered During sta...

Page 11: ...output contact The digital output contact status can be changed by the user only by sending the proper SMS to the interface The below table shows the digital output LED indication depending on the di...

Page 12: ...n through serial cable see Local Configuration section To proceed with the remote configuration select the Remote connection type on HI504901SW by checking the proper box On HI 504901 phone number fie...

Page 13: ...each update with a wrong PIN value an attempt will be lost The SMS s stored in the SIM card will be erased Note If an error occurs on cellular phone before pressing the Up date All button the cellula...

Page 14: ...card is used If the feature is enabled it is necessary to insert a value to proceed with the configuration On Expiration Date field it is possible to set the SIM card expiration date dd mm yy this is...

Page 15: ...ed and this operation lasts about 30 seconds the cel lular phone will be turned off and on again to verify the PIN code correctness 15 days 7 days and 1 day before the expiration date warn ing SMS mes...

Page 16: ...ge Each instrument must have a different value for this field Note If the SIM card PIN number is longer than 4 digits it has to be changed before using it in the HI504901 interface To do this insert t...

Page 17: ...axi mum length of this description is 30 characters Digital output section It is possible to enable or disable the control of the digital output status connected to the HI504901 by checking the proper...

Page 18: ...for instruments connected to the RS232 ports will be INSTRUMENTNAME RS232 NUMBEROFTHERS232 CHANNEL Alarms ALARM SPECIFIC TO THE INSTRU MENT Rem msg NUMBER OF REMAINING MESSAGES The format of alarm me...

Page 19: ...turned off or only active alarms when all active alarms can not be sent within one SMS the maximum number of characters for an SMS is 160 more than one message will be sent each one having as header t...

Page 20: ...to send while SMS de lay item sets the delay time between two subsequent alarm messages INFORMATION SMSs In addition to the alarm messages already described the cellu larphoneinterfacecansendinformat...

Page 21: ...01 or HI8002 controller a third SMS will be sent with the label INFO3 3 This message will contain the got from the controlled instruments and could be requested for example after an alarm message rega...

Page 22: ...ne number s This particular situation is managed as an error occurrence and a confir mation is waited When this happens no more messages will be sent by the interface until the error is disactivated l...

Page 23: ...or indication is given by interface green and red LEDs disable enable blink together To disactivate this error and restore the SMS service it is necessary to re charge the SIM card credit Note Every t...

Page 24: ...messages are temporarily sus pended PC COMMUNICATION HI504901 interface can act as a gateway for a connection through a remote computer With this feature a seamless data connection will be established...

Page 25: ...nd PWD interface password CR and the interface will answer ACK char 0x06 if the password is correct or CAN char 0x18 if password is not correct or NAK char 0x15 if command syntax is wrong and hangs up...

Page 26: ...R has to be sent to cellular phone interface For example if target device is attached on RS232 1 port and its communication rate is 2400 bps then the command C0B CR must be issued The cellular phone i...

Page 27: ...rface configuration and PC communications both lo cal remote are not allowed until mains supply comes back When the mains power supply comes back the cellular phone turns on and cellular initializatio...

Page 28: ...capability make sure that the HI504901 is turned off Note Even if a battery problem persists all HI504901 functions are still available FAULT CONDITIONS SELFTEST PROCEDURES HI504901 is provided with s...

Page 29: ...LED blinking fast cellular engine registered to the net work and a call is in progress LED INDICATORS Find here below a brief explanation of all possible LEDs indi cations of HI504901 interface RS232...

Page 30: ...Only two LEDs green and red LEDs for SMS sending enable disable are available to give information to the user about active errors and no more than one can be shown at the same time For this reason ev...

Page 31: ...chargeorsubstitutetheSIMcard Change the SIM card with one enabled for voice calls Note Some SIM cards can be used both for data and voice calls Please check the networkoperatorforfurtherdetails During...

Page 32: ...with remote device RunHI504901SWapplication software and change the options for SMS sending logic on digitalinput window EnableSMSsendingthroughexternalswitch Can not control digital output contact t...

Page 33: ...d electrical shock do not use this instrument when voltage at the measurement surface exceed 24VAC or 60VDC To avoid damage or burns do not perform any measurement in microwave ovens External cables t...

Page 34: ...67 66 USER NOTES...

Page 35: ...584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 South Africa Tel 011 615 607...

Reviews: