background image

31

30

• “pH Range alarm”: this alarm is triggered when the real

time value of the pH sensor #1 is outside a dead band set
by the user during instrument setup.

• “EC Range alarm”: this alarm is triggered when the real

time value of the EC sensor #1 is outside a dead band set
by the user during instrument setup.

• “#n Filter dirty alarm”: this alarm indicates that the “n”-th

filter requires cleaning.

• “#n Fertilizer low level alarm”: this alarm is triggered if the

fertilizer level of the “n”-th designated tank reached a mini-
mum value.

• “Acid low level alarm”: this alarm is triggered if the level of

the acid (or alkaline) tank reached a minimum value.

• “No water supply alarm”: this alarm is triggered due to an

incoming water supply failure event.

Please refer to the 

HI8001

 and 

HI8002 

Instruction Manual

for a complete explanation of the above alarm conditions.

RS485 OPTIONS SETTING

In this section it is possible to set all options related to the
RS485 communication channel towards the instruments with
an RS485 output.
The maximum number of instruments that can be connected to
the same bus is 32. To set the instrument options, click on one
of the 32 available buttons and then set the following features:
• Instrument name: select the name of the instrument con-

nected to the RS485 from a list (

HI700

HI710

HI23

,

HI24

HI504

 and 

HI504910

). To disable an instrument

previously inserted, select “NO INSTRUMENT”.

• Baud rate: once the instrument has been selected, the de-

fault baud rate value is automatically proposed. It is pos-
sible to select a different value from a list (1200, 2400,
4800 and 9600 bps).

Note

All instruments connected to RS485 must be set with the same
baud rate value. Once a new instrument is added in

HI504901

 configuration, the current RS485 baud rate value

is proposed: it is possible to change this value, but this modi-
fication will be applied to all the instruments previously con-
figured.

• RS485 address: it is the RS485 address of the instrument

(allowed values are within 0 to 99 range). Each instrument
must have a different value for this field.

Note

If the SIM card PIN number is longer than 4 digits, it has to
be changed before using it in the

 HI504901

 interface. To

do this, insert the SIM card in a cellular phone and change
the old PIN number with a new one with value from 0000 to
9999, while paying attention to remember the new code (please
refer to the cellular phone instruction manual for the correct
procedure). After that, it is possible to insert the SIM card in
the 

HI504901

 (see “Installation” section) and use the new

PIN number in the next interface local configuration.

RS232 OPTIONS SETTING

In this section it is possible to set all options related to the
RS232 communication channels towards the instruments within
an RS232 output port. A channel can be enabled only if an
instrument is connected to it. For each channel the following
options have to be set:
• Instrument name: select the name of the instrument con-

nected to the RS232 channel from a list (

HI8001

 and

HI8002

).

• Baud rate: once the instrument has been selected, the de-

fault baud rate value is automatically proposed (i.e. 9600
bps). It is possible to select a different value from a list
(1200, 2400, 4800 and 9600 bps).

• Password: click on the “Password” button to set the correct

instrument password. Insert the password (masked display)
and then confirm it. If no password is set in 

HI8001

 and

HI8002

 controllers, use the default password (i.e. 0000).

• Messages options: to set the alarm message options, click on

the “Message Options” button. The available options de-
pend on the selected instrument (see also following sections).

HI8001 and HI8002 Message Options:

In this window it is possible to choose in which alarm condi-
tions for 

HI8001 

and 

HI8002

 controllers an SMS will be sent

to the selected phone number(s).
Available alarm conditions for 

HI8001 

and

 HI 8002

 are:

• “pH Sensor alarm”: this alarm is triggered if the difference

between the real time values of the two primary pH sensors
exceed the specified maximum allowable difference.

• “EC Sensor alarm”: this alarm is triggered if the difference

between the real time values of the two primary EC sensors
exceed the specified maximum allowable difference.

Summary of Contents for HI 504901

Page 1: ...HI 504901 GSM Supervisor Instruction Manual...

Page 2: ...tput DIG OUT 13 RS485 Serial Port 13 Power Supply Input 15 Battery 15 SIM Card Insertion 16 START UP 18 IDLE MODE 19 INTERFACE CONFIGURATION 22 Application Software Installation 22 Starting Configurat...

Page 3: ...failure If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization...

Page 4: ...battery which provides power supply in case of mains power failure PRELIMINARY EXAMINATION Remove the instrument from the packing material and exam ine it carefully to make sure that no damage has occ...

Page 5: ...sealed lead acid 12V 0 8Ah 2 On board battery connector 3 GSM module 4 LED indicators see figure on next page 5 RS232 auxiliary communication port for local PC communication only 6 RS232 channel 1 se...

Page 6: ...ity of electronic devices Unplug and turn off the unit in hospitals or near medical devices like pacemakers or hearing aids The module may interfere with the operation of these devices Unplug and turn...

Page 7: ...is standard suitable for data transmission in noisy environ ments INSTALLATION RS232 AUXILIARY PORT HI504901 is provided with an auxiliary RS232 port D Sub 9 pole female plug for connection to a local...

Page 8: ...p to 1 2 Km typical with 24 AWG cable Loads up to 32 typical Internal termination none Connections The connections for the provided 2 pin RS485 terminal are as shown on the below diagram The instrumen...

Page 9: ...10 hours before turning ON the power switch In this manner the new battery can be charged SIM CARD INSERTION Before using the HI504901 insert a SIM card Press the SIM card holder ejector on the cellul...

Page 10: ...ignal2 green ON START UP Turn ON the power switch 14 on page 8 the red SMS disable LED lights up and will stay in this state until the start up procedure ended and the idle mode is en tered During sta...

Page 11: ...output contact The digital output contact status can be changed by the user only by sending the proper SMS to the interface The below table shows the digital output LED indication depending on the di...

Page 12: ...n through serial cable see Local Configuration section To proceed with the remote configuration select the Remote connection type on HI504901SW by checking the proper box On HI 504901 phone number fie...

Page 13: ...each update with a wrong PIN value an attempt will be lost The SMS s stored in the SIM card will be erased Note If an error occurs on cellular phone before pressing the Up date All button the cellula...

Page 14: ...card is used If the feature is enabled it is necessary to insert a value to proceed with the configuration On Expiration Date field it is possible to set the SIM card expiration date dd mm yy this is...

Page 15: ...ed and this operation lasts about 30 seconds the cel lular phone will be turned off and on again to verify the PIN code correctness 15 days 7 days and 1 day before the expiration date warn ing SMS mes...

Page 16: ...ge Each instrument must have a different value for this field Note If the SIM card PIN number is longer than 4 digits it has to be changed before using it in the HI504901 interface To do this insert t...

Page 17: ...axi mum length of this description is 30 characters Digital output section It is possible to enable or disable the control of the digital output status connected to the HI504901 by checking the proper...

Page 18: ...for instruments connected to the RS232 ports will be INSTRUMENTNAME RS232 NUMBEROFTHERS232 CHANNEL Alarms ALARM SPECIFIC TO THE INSTRU MENT Rem msg NUMBER OF REMAINING MESSAGES The format of alarm me...

Page 19: ...turned off or only active alarms when all active alarms can not be sent within one SMS the maximum number of characters for an SMS is 160 more than one message will be sent each one having as header t...

Page 20: ...to send while SMS de lay item sets the delay time between two subsequent alarm messages INFORMATION SMSs In addition to the alarm messages already described the cellu larphoneinterfacecansendinformat...

Page 21: ...01 or HI8002 controller a third SMS will be sent with the label INFO3 3 This message will contain the got from the controlled instruments and could be requested for example after an alarm message rega...

Page 22: ...ne number s This particular situation is managed as an error occurrence and a confir mation is waited When this happens no more messages will be sent by the interface until the error is disactivated l...

Page 23: ...or indication is given by interface green and red LEDs disable enable blink together To disactivate this error and restore the SMS service it is necessary to re charge the SIM card credit Note Every t...

Page 24: ...messages are temporarily sus pended PC COMMUNICATION HI504901 interface can act as a gateway for a connection through a remote computer With this feature a seamless data connection will be established...

Page 25: ...nd PWD interface password CR and the interface will answer ACK char 0x06 if the password is correct or CAN char 0x18 if password is not correct or NAK char 0x15 if command syntax is wrong and hangs up...

Page 26: ...R has to be sent to cellular phone interface For example if target device is attached on RS232 1 port and its communication rate is 2400 bps then the command C0B CR must be issued The cellular phone i...

Page 27: ...rface configuration and PC communications both lo cal remote are not allowed until mains supply comes back When the mains power supply comes back the cellular phone turns on and cellular initializatio...

Page 28: ...capability make sure that the HI504901 is turned off Note Even if a battery problem persists all HI504901 functions are still available FAULT CONDITIONS SELFTEST PROCEDURES HI504901 is provided with s...

Page 29: ...LED blinking fast cellular engine registered to the net work and a call is in progress LED INDICATORS Find here below a brief explanation of all possible LEDs indi cations of HI504901 interface RS232...

Page 30: ...Only two LEDs green and red LEDs for SMS sending enable disable are available to give information to the user about active errors and no more than one can be shown at the same time For this reason ev...

Page 31: ...chargeorsubstitutetheSIMcard Change the SIM card with one enabled for voice calls Note Some SIM cards can be used both for data and voice calls Please check the networkoperatorforfurtherdetails During...

Page 32: ...with remote device RunHI504901SWapplication software and change the options for SMS sending logic on digitalinput window EnableSMSsendingthroughexternalswitch Can not control digital output contact t...

Page 33: ...d electrical shock do not use this instrument when voltage at the measurement surface exceed 24VAC or 60VDC To avoid damage or burns do not perform any measurement in microwave ovens External cables t...

Page 34: ...67 66 USER NOTES...

Page 35: ...584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 South Africa Tel 011 615 607...

Reviews: