POWER SUPPLY INPUT

Use a 12VDC adapter (

HI710005

 or

HI710006

) and connect the black wire to

the positive input terminal (marked “+”) and
the black & white wire to the negative one
(marked “-”).

All hard wiring can be accomplished through five wa-
tertight cable glands on the bottom side of the case,
by passing wires through the rubber grommets and
tightening the nuts.

Note

All connected cables should end with cable lugs.

BATTERY

FOR SAVING BATTERY, THE HI504901 IS SUP-
PLIED WITH BATTERY NOT CONNECTED.
PLUG THE BATTERY CONNECTOR BEFORE
STARTING OPERATION.

Use only rechargeable sealed lead-acid battery with
12 V and 0.8 Ah capability.

To substitute battery, please follow below instructions:

1. turn off the 

HI504901

: turn OFF the power switch and

disconnect the interface from the mains;

2. remove the screws on the bottom of the two battery clips;
3. disconnect the battery cable from its connector;

4. substitute the old battery with a new one;

5. fix back the two girdles by tightening the screws;

Specifications:

The RS485 standard is implemented in 

HI504901

 interface

with the following characteristic:
Data rate:

up to 9600 (selected while configuring the

interface through 

HI504901SW

 software)

Communication:

bidirectional Half-Duplex

Line length:

up to 1.2 Km typical, with 24 AWG cable

Loads:

up to 32 typical

Internal termination: none

Connections:

The connections for the provided 2-pin RS485 terminal are
as shown on the below diagram.
The instrument has no internal line termination: to terminate
the line an external resistor equal to the characteristic line im-
pedance (typically 120 

) must be added at both line ends.

HI504901

 RS485 port can connect up to 32 instruments on

the same physical network, with a total line length up to 1.2
Km using a 24 AWG cable.

All the instruments connected to the RS485 port are “slave”
devices that are queried by the 

HI504901

 supervisor. In

other words, the interface can work only as a “master” com-
ponent, whereas the connected instruments work as remote
terminal equipments answering to the commands only.
Each instrument (except the 

HI504901

 interface itself) is iden-

tified by its Process ID number, included within the 00 to 99
interval, which corresponds to the Process ID configured
through the proper setup item (please refer to the instrument
instruction manual for a complete explanation).

Note      

If an instrument does not recognize the address within the

command string, then it discard all the following bytes.

As additional feature, the 

HI504901

 interface is also pro-

vided with internal Fail Safe Open Line protection method.

To minimize electromagnetic interferences, it is recommended
to use shielded and twisted pair cable to connect the units.

Summary of Contents for HI 504901

Page 1: ...HI 504901 GSM Supervisor Instruction Manual...

Page 2: ...tput DIG OUT 13 RS485 Serial Port 13 Power Supply Input 15 Battery 15 SIM Card Insertion 16 START UP 18 IDLE MODE 19 INTERFACE CONFIGURATION 22 Application Software Installation 22 Starting Configurat...

Page 3: ...failure If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization...

Page 4: ...battery which provides power supply in case of mains power failure PRELIMINARY EXAMINATION Remove the instrument from the packing material and exam ine it carefully to make sure that no damage has occ...

Page 5: ...sealed lead acid 12V 0 8Ah 2 On board battery connector 3 GSM module 4 LED indicators see figure on next page 5 RS232 auxiliary communication port for local PC communication only 6 RS232 channel 1 se...

Page 6: ...ity of electronic devices Unplug and turn off the unit in hospitals or near medical devices like pacemakers or hearing aids The module may interfere with the operation of these devices Unplug and turn...

Page 7: ...is standard suitable for data transmission in noisy environ ments INSTALLATION RS232 AUXILIARY PORT HI504901 is provided with an auxiliary RS232 port D Sub 9 pole female plug for connection to a local...

Page 8: ...p to 1 2 Km typical with 24 AWG cable Loads up to 32 typical Internal termination none Connections The connections for the provided 2 pin RS485 terminal are as shown on the below diagram The instrumen...

Page 9: ...10 hours before turning ON the power switch In this manner the new battery can be charged SIM CARD INSERTION Before using the HI504901 insert a SIM card Press the SIM card holder ejector on the cellul...

Page 10: ...ignal2 green ON START UP Turn ON the power switch 14 on page 8 the red SMS disable LED lights up and will stay in this state until the start up procedure ended and the idle mode is en tered During sta...

Page 11: ...output contact The digital output contact status can be changed by the user only by sending the proper SMS to the interface The below table shows the digital output LED indication depending on the di...

Page 12: ...n through serial cable see Local Configuration section To proceed with the remote configuration select the Remote connection type on HI504901SW by checking the proper box On HI 504901 phone number fie...

Page 13: ...each update with a wrong PIN value an attempt will be lost The SMS s stored in the SIM card will be erased Note If an error occurs on cellular phone before pressing the Up date All button the cellula...

Page 14: ...card is used If the feature is enabled it is necessary to insert a value to proceed with the configuration On Expiration Date field it is possible to set the SIM card expiration date dd mm yy this is...

Page 15: ...ed and this operation lasts about 30 seconds the cel lular phone will be turned off and on again to verify the PIN code correctness 15 days 7 days and 1 day before the expiration date warn ing SMS mes...

Page 16: ...ge Each instrument must have a different value for this field Note If the SIM card PIN number is longer than 4 digits it has to be changed before using it in the HI504901 interface To do this insert t...

Page 17: ...axi mum length of this description is 30 characters Digital output section It is possible to enable or disable the control of the digital output status connected to the HI504901 by checking the proper...

Page 18: ...for instruments connected to the RS232 ports will be INSTRUMENTNAME RS232 NUMBEROFTHERS232 CHANNEL Alarms ALARM SPECIFIC TO THE INSTRU MENT Rem msg NUMBER OF REMAINING MESSAGES The format of alarm me...

Page 19: ...turned off or only active alarms when all active alarms can not be sent within one SMS the maximum number of characters for an SMS is 160 more than one message will be sent each one having as header t...

Page 20: ...to send while SMS de lay item sets the delay time between two subsequent alarm messages INFORMATION SMSs In addition to the alarm messages already described the cellu larphoneinterfacecansendinformat...

Page 21: ...01 or HI8002 controller a third SMS will be sent with the label INFO3 3 This message will contain the got from the controlled instruments and could be requested for example after an alarm message rega...

Page 22: ...ne number s This particular situation is managed as an error occurrence and a confir mation is waited When this happens no more messages will be sent by the interface until the error is disactivated l...

Page 23: ...or indication is given by interface green and red LEDs disable enable blink together To disactivate this error and restore the SMS service it is necessary to re charge the SIM card credit Note Every t...

Page 24: ...messages are temporarily sus pended PC COMMUNICATION HI504901 interface can act as a gateway for a connection through a remote computer With this feature a seamless data connection will be established...

Page 25: ...nd PWD interface password CR and the interface will answer ACK char 0x06 if the password is correct or CAN char 0x18 if password is not correct or NAK char 0x15 if command syntax is wrong and hangs up...

Page 26: ...R has to be sent to cellular phone interface For example if target device is attached on RS232 1 port and its communication rate is 2400 bps then the command C0B CR must be issued The cellular phone i...

Page 27: ...rface configuration and PC communications both lo cal remote are not allowed until mains supply comes back When the mains power supply comes back the cellular phone turns on and cellular initializatio...

Page 28: ...capability make sure that the HI504901 is turned off Note Even if a battery problem persists all HI504901 functions are still available FAULT CONDITIONS SELFTEST PROCEDURES HI504901 is provided with s...

Page 29: ...LED blinking fast cellular engine registered to the net work and a call is in progress LED INDICATORS Find here below a brief explanation of all possible LEDs indi cations of HI504901 interface RS232...

Page 30: ...Only two LEDs green and red LEDs for SMS sending enable disable are available to give information to the user about active errors and no more than one can be shown at the same time For this reason ev...

Page 31: ...chargeorsubstitutetheSIMcard Change the SIM card with one enabled for voice calls Note Some SIM cards can be used both for data and voice calls Please check the networkoperatorforfurtherdetails During...

Page 32: ...with remote device RunHI504901SWapplication software and change the options for SMS sending logic on digitalinput window EnableSMSsendingthroughexternalswitch Can not control digital output contact t...

Page 33: ...d electrical shock do not use this instrument when voltage at the measurement surface exceed 24VAC or 60VDC To avoid damage or burns do not perform any measurement in microwave ovens External cables t...

Page 34: ...67 66 USER NOTES...

Page 35: ...584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 South Africa Tel 011 615 607...

Reviews: