4. Installation
Please check first that the oscilloscope corresponds to the SAFETY CLASS I conditions
(three-conductor power cord with protective earthing contacts on plugs and oscilloscope).
If not, it is not permitted to install the interface. In this case contact your
HAMEG
distributor.
4.1.
Switch off the oscilloscope and disconnect the power connector.
4.2.
Place the oscilloscope on a soft surface with the front side facing down.
4.3.
Remove the cap nuts underneath the power connector socket and on the right
side of the back cover.
Attention!
If the oscilloscope is not supplied with a cap nut underneath the power
connector socket, it must additionally be attached. The required mechanical
parts are included. Then the following steps must be performed first.
4.3.1
Remove the cap nut on the left side of the back cover and pull it off.
4.3.2
Carefully pull the case off towards the back.
4.3.3
Remove the right back cover mounting screw and replace it by the
supplied longer (14mm) screw.
4.3.4
As before again place the scope on a soft surface and carefully slide
the case on.
4.3.5
Slide the leaf spring over the thread of the supplied screw, with the angle
facing to the screw head and thereafter a lock washer.
4.3.6
Slide the screw from the inner side of the back cover through the 4mm
diameter hole (underneath the power connector socket).
4.3.7
Slide a lock washer over the thread and mount the screw with the
supplied nut. The leaf spring must be in vertical position facing the
power connector socket.
4.3.8
Replace the back cover and mount it with the original cap nut on the left
side of the back cover.
4.4.
Place the interface in such a way on the back cover that the 9pole D-Sub male
connector of the interface is located above the 9pole D-Sub female connector of
the scope and press the interface against the back cover.
4.5.
Attach the interface with the formerly removed cap nuts to have a proper electrical
connection to the interface. This is required to meet the EMC regulations.
Please note:
The interface cabinet and the interface connectors are galvanically connected
to the protective earth (Safety Class I) via the oscilloscope.