Hameg HM 1004-3 Series Manual Download Page 9

9

Subject to change without notice

Type of signal voltage

H

between 0.5 and 8div, if possible 3.2 to 8div,

U

between 0.5mVpp and 160Vpp,

D

between 1mV/div and 20V/div in 1-2-5 sequence.

Examples:

Set deflection coefficient 

D

 = 50mV/div   0.05V/div,

observed display height 

H

 = 4.6div,

required voltage 

U

 = 0.05x4.6 = 0.23Vpp.

Input voltage

 U

 = 5Vpp,

set deflection coefficient 

D

 = 1V/div,

required display height 

H

 = 5:1 = 5div.

Signal voltage 

U

 = 230Vrmsx2

2 = 651Vpp

(voltage > 160Vpp, with probe 10:1: 

U

 = 65.1Vpp),

desired display height 

H

 = min. 3.2div, max. 8div,

max. deflection coefficient 

D

 = 65.1:3.2 = 20.3V/div,

min. deflection coefficient 

D

 = 65.1:8 = 8.1V/div,

adjusted deflection coefficient 

D

 = 10V/div.

The previous examples are related to the crt graticule reading.
The results can also be determined with the aid of the 

V

cursor measurement (

please note “controls and readout”

).

The input voltage must not exceed 400V, independent from
the polarity.

If an AC voltage which is superimposed on a DC voltage is
applied, the maximum peak value of both voltages must not
or -400V. So for AC voltages with a mean value of
zero volt the maximum peak to peak value is 800Vpp.

If attenuator probes with higher limits are used, the
probes limits are valid only if the oscilloscope is set to
DC input coupling.

If DC voltages are applied under AC input coupling conditions
the oscilloscope maximum input voltage value remains 400V.
The attenuator consists of a resistor in the probe and the 1M

input resistor of the oscilloscope, which are disabled by the
AC input coupling capacity when AC coupling is selected. This
also applies to DC voltages with superimposed AC voltages.

It also must be noted that due to the capacitive resistance of
the AC input coupling capacitor, the attenuation ratio depends
on the signal frequency. For sinewave signals with frequen-
cies higher than 40Hz this influence is negligible.

With the above listed exceptions

 HAMEG 

10:1 probes can be

used for DC measurements up to 600V or AC voltages (with
a mean value of zero volt) of 1200Vpp. The 100:1 probe

 HZ53

allows for 1200V DC or 2400Vpp for AC.

It should be noted that its AC peak value is derated at higher
frequencies. If a normal x10 probe is used to measure high
voltages there is the risk that the compensation trimmer
bridging the attenuator series resistor will break down caus-
ing damage to the input of the oscilloscope.

However, if for example only the residual ripple of a high
voltage is to be displayed on the oscilloscope, a normal x10
probe is sufficient. In this case, an appropriate high voltage
capacitor (approx. 22-68nF) must be connected in series with
the input tip of the probe. With 

Y-POS.

 control (input coupling

to 

GD

) it is possible to use a 

horizontal graticule line as

reference line for ground potential before the measure-
ment.

 It can lie below or above the horizontal central line

according to whether positive and/or negative deviations
from the ground potential are to be measured.

Total value of input voltage

The dotted line shows a voltage alternating at zero volt level.
If superimposed on a DC voltage, the addition of the positive
peak and the DC voltage results in the max. voltage (DC +
ACpeak).

Time Measurements

As a rule, most signals to be displayed are periodically
repeating processes, also called periods. The number of
periods per second is the repetition frequency. Depending on
the timebase setting (

TIME/DIV.

-knob) indicated by the

readout, one or several signal periods or only a part of a period
can be displayed. The time coefficients are stated in 

ms/div

,

µs/div

 or 

ns/div

. The following examples are related to the

crt graticule reading. The results can also be determined with
the aid of the 

T and 1/

T cursor measurement (

please note

“ Controls and Readout”

).

The duration of a signal period or a part of it is determined by
multiplying the relevant time (horizontal distance in div) by the
(calibrated) time coefficient displayed in the readout .
Uncalibrated, the timebase speed can be reduced until a
maximum factor of 2.5 is reached. Therefore any intermedi-
ate value is possible within the 1-2-5 sequence.

With the designations

L

= displayed wave length in div of one period,

T

= time in seconds for one period,

F

= recurrence frequency in Hz of the signal,

Tc

= time coefficient in ms, µs or ns/div and the relation

F

= 1/T, the following equations can be stated:

However, these four values are not freely selectable. They
have to be within the following limits:

L

between 0.2 and 10div, if possible 4 to 10div,

T

between 5ns and 5s,

F

between 0.5Hz and 100MHz,

Tc

between 50ns/div and 500ms/div in 1-2-5 sequence
(with X-MAG. (x10) inactive), and

Tc

between 5ns/div and 50ms/div in 1-2-5 sequence
(with X-MAG. (x10) active).

Examples:

Displayed wavelength L = 7div,
set time coefficient Tc = 100ns/div,
required period T = 7x100x10-9 = 0.7µs
required rec. freq. F = 1:(0.7x10-6) = 1.428MHz.
Signal period T = 1s,

Summary of Contents for HM 1004-3 Series

Page 1: ...Instruments Oscilloscope HM 1004 3 01 02 03 ENGLISH MANUAL HANDBUCH MANUEL...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...res 23 Phase difference measurement in DUAL mode Yt 24 Phase difference measurement in DUAL mode 24 Measurement of amplitude modulation 24 Triggering and timebase 25 Automatic Peak value Triggering 25...

Page 4: ...ollowing hints and conditions regarding emission and immunity should be observed 1 Data cables For the connection between instruments resp their interfaces and external devices computer printer etc su...

Page 5: ...x ambient temperature 0 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Color techno brown Cabinet W 285 H 125 D 380 mm Subject to change without notice 08 00 This microprocessor...

Page 6: ...f an extension cord without a protective conductor The mains line plug must be inserted before connec tions are made to measuring circuits The grounded accessible metal parts case sockets jacks and th...

Page 7: ...helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain intervals Only in this way is it largel...

Page 8: ...it a true signal display The display of sinusoidal signals within the bandwidth limits causes no prob lems but an increasing error in measurement due to gain reduction must be taken into account when...

Page 9: ...caus ing damage to the input of the oscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case...

Page 10: ...asurement of the rise or fall time is not limited to the trace dimensions shown in the above diagram It is only particularly simple in this way In principle it is possible to measure in any display po...

Page 11: ...Hz There fore the derating curve of the attenuator probe type con cerned must be taken into account The selection of the ground point on the test object is important when displaying small signal volta...

Page 12: ...iver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile memories These can be used by the ope...

Page 13: ...with this control knob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a h...

Page 14: ...aid of the X POS control As the X expansion results in a higher timebase speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over Please note...

Page 15: ...hannel II followed by XY Except the cursor lines which may be active all other readout information including the trigger point symbol are switched off In addition to all trigger and timebase related c...

Page 16: ...l switched off NR high frequency noise rejected LF low pass filter cuts off frequencies above approx 1 5kHz TVL TV signal line pulse triggering trigger point symbol switched off TVF TV signal frame pu...

Page 17: ...e TIME DIV 24 control knob is operative only for this timebase The readout then only displays the A time coefficient The timebase settings for this condition are stored if the timebase mode is changed...

Page 18: ...switch and B vernier VAR LED on In the latter case the TIME DIV knob can be used in the same way as de scribed before under A timebase condition Underneath the front panel sector described above the...

Page 19: ...r related con trols assumes that the readout is visible and the component tester is switched off 35 MENU Pushbutton Pressing and holding the pushbutton activates the display of the MAIN MENU It contai...

Page 20: ...as one signal sum or difference As the result can only be determined if both calibrated deflection coefficients are equal the CHI II 36 selection function is deactivated In that case the readout indic...

Page 21: ...ted to the oscilloscope input s Check that the device under test is switched off and connect the test lead s to the test point s Then switch on the instrument and afterwards the device under test The...

Page 22: ...rm distortion Prerequisite for this HF compensation is a square wave generator with fast risetime typically 4ns and low output impedance approx 50 providing 0 2V at a frequency of approx 1MHz The cali...

Page 23: ...this operation mode is the DUAL XY 18 pushbutton please note section Con trols and Readout In XY mode the timebase is deactivated The signal applied to the input of channel II front panel marking HOR...

Page 24: ...ng for one period in div In the example illustrated t 3div and T 10div The phase difference in degrees is calculated from Relatively small phase angles at lower frequencies can be measured more accura...

Page 25: ...ode for the entry into difficult measuring problems e g when the test signal is unknown relating to amplitude frequency or shape Presetting of all parameters is now possible with automatic triggering...

Page 26: ...5div height The polarity of the synchronization pulse is critical for the slope selection If the displayed sync pulses are above the picture field contents leading edge positive going then the slope...

Page 27: ...vated As the external trigger signal applied at the TRIG EXT socket normally has no relation to the signal height of the displayed signal the trigger point symbol is switched off The external trigger...

Page 28: ...nt This allows the sweep to begin on any portion of a signal except the first millimeters If the display shows jitter it is possible to select triggering of the B timebase after the elapsed delay time...

Page 29: ...lugs and with test prod alligator clip or sprung hook are required The test leads are connected as described in section Con trols and Readout Test Procedure Caution Do not test any component in live c...

Page 30: ...r a transistor the figures b e and b c are important The figure e c can vary but a vertical line only shows short circuit condition These transistor test patterns are valid in most cases but there are...

Page 31: ...devices In case of disregard of the safety warnings contained in this manual HAMEG refuses any liability regarding personal injury and or damage of equip ment Operation The oscilloscope is supplied w...

Page 32: ...Subject to change without notice 32 Front Panel HM1004 3...

Page 33: ...33 Subject to change without notice...

Page 34: ...Subject to change without notice 34...

Page 35: ...MANUAL HANDBUCH MANUEL...

Page 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: