Hameg HM 1004-3 Series Manual Download Page 20

Subject to change without notice

20

Controls and Readout

between different deflection coefficients. Consequently
the pushbutton is inactivated in combination with 

V

measurements.

1/

∆∆∆∆∆

t:

If the 

V function is not present, briefly pressing the

button selects between time and frequency measure-
ment which is not available in XY mode. The vertical
cursor lines and the measurement result apply to the
active timebase (A or B, resp. B in alternate timebase
mode).
In calibrated timebase condition the readout displays

∆∆∆∆∆

t:...” 

if time measurement is chosen. After switching

over to 1/

t (frequency) “f:...” is displayed. If the timebase

is uncalibrated the readout displays

 “

∆∆∆∆∆

t>...” 

or

 “f<...”

.

(37) TRK 

- CURSOR-Track function.

Briefly pressing simultaneously both buttons

 ON/OFF 

-

CHI/II  - 1/

∆∆∆∆∆

t (36) 

and

 I/II - 

∆∆∆∆∆

V/

∆∆∆∆∆

t (38) 

switches over from

single cursor line operation to track mode and vice versa.

In TRK (track) mode both cursor lines are indicated as
active (both dotted lines uninterrupted).

(38) I/II - 

∆∆∆∆∆

V/

∆∆∆∆∆

- Pushbutton with two functions.

I/II:

Briefly pressing this button changes the active (controlla-
ble) cursor in the sequence I - II - I, if TRK (track) mode is
not active. The active cursor is indicated by a continuously
dotted line. An interrupted dotted line indicates the
inactive cursor.

∆∆∆∆∆

V/

∆∆∆∆∆

t:

Pressing and holding this pushbutton changes from volt-
age to time (or frequency) measurement and vice versa.
In XY mode the instrument is automatically set to 

V.

∆∆∆∆∆

V:

Please note!
In all 

∆∆∆∆∆

V (voltage) measurement conditions, the divi-

sion ratio of the probe(s) must be taken into account.
The voltage value displayed in the readout must be
multiplied by 100 if e.g. a x100 (100:1) probe is used. In
case of x10 (10:1) probes, the probe factor can be
automatically included (see item (28) and (32)).

item 1: Timebase mode.

(CH I or CH II single channel mode, DUAL and ADD)

In the voltage measurement condition, the cursors are
displayed as horizontal lines and the result is displayed in
the readout.

Single channel mode (channel I or channel II).

The 

V measuring result is automatically related to the

deflection coefficient of the active channel. The readout
displays

  “

∆∆∆∆∆

V1...” 

or

  “

∆∆∆∆∆

V2...”

.

Dual mode.

The CURSOR lines must be set on the channel I or
channel II signal. As the deflection coefficients may be

different, it will be required to select between the deflec-
tion coefficients of channel I and II.

 Please note item (36)

CHI/II

.

Addition (ADD) mode.

In

 ADD 

(addition) mode normally two input signals are

displayed as one signal (sum or difference). As the result
can only be determined if both (calibrated) deflection
coefficients are equal, the

 CHI/II (36) 

selection function

is deactivated. In that case the readout indicates

 “

∆∆∆∆∆

V...”

without any additional channel information. Different
deflection coefficient settings or uncalibrated deflection
coefficient(s) are indicated in the readout as

 “Y1<>Y2”

.

item 2: XY mode.

In XY mode the instrument is automatically set to

 

∆∆∆∆∆

V

measurement. The deflection coefficient selected for
each channel may be different, thus as in DUAL mode the

∆∆∆∆∆

V

 cursor measurement requires a channel selection.

Under channel I (Y signal) measuring condition the cursor
lines are displayed as horizontal lines and the readout
displays

  “

∆∆∆∆∆

VY...”

. The cursor lines are displayed as

vertical lines and the readout indicates

  “

∆∆∆∆∆

VX...” 

if the

channel II deflection coefficient is chosen.

∆∆∆∆∆

t:

In time or frequency measurement condition two vertical
cursor lines are displayed.  The measurement result is
indicated as

  “

∆∆∆∆∆

t...” 

(time measurement) or

  “f...” 

(fre-

quency measurement).

 Please note item (36) 1/

∆∆∆∆∆

t

.

NOTE:
For frequency measurement, the distance between
the cursors must equal exactly one signal period.
In XY mode the timebase is switched off. Conse-
quently time or frequency measurement is disabled.

(39)CURSOR 

- Center biased lever.

The active cursor line (in track mode: both lines) can be
shifted in the required direction, until the graticule limits
are reached. The directions are marked on the front panel
and depend on the selected measurement (

U, 

t or 1/

t). Which of two shift speeds is used, depends on how

far the

 CURSOR 

knob is pressed. With slight knob

pressure the cursor line(s) moves slowly. If the knob is
pressed to the full extent the cursor moves fast. If the
knob is released it automatically reverts to the center
position and the cursor shift stops.

(40)CAL. 

- Pushbutton and concentric socket.

A squarewave signal of 0.2Vpp ±1% is available from the
socket for probe adjustment purposes. The signal fre-
quency depends on the pushbutton setting. If the
pushbutton is released it is approx. 1kHz and can be
switched over (depressed) to approx. 1MHz. The pulse
duty factor may deviate from 1:1 and the frequency is not
calibrated.

(41)CT 

- Pushbutton and banana jack.

Pressing the pushbutton switches the instrument over
from oscilloscope to component test mode and vice
versa.
In component test mode, all controls are deactivated with
the exception of the

 CT 

button,

 AUTO SET (2) 

and

INTENS (4). 

All LEDs except

  “A” 

or

  “RO” 

(associated

with the INTENS knob) are dark. The readout displays only

“CT” 

in this condition.

One test lead is connected to the CT socket. The second

Summary of Contents for HM 1004-3 Series

Page 1: ...Instruments Oscilloscope HM 1004 3 01 02 03 ENGLISH MANUAL HANDBUCH MANUEL...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...res 23 Phase difference measurement in DUAL mode Yt 24 Phase difference measurement in DUAL mode 24 Measurement of amplitude modulation 24 Triggering and timebase 25 Automatic Peak value Triggering 25...

Page 4: ...ollowing hints and conditions regarding emission and immunity should be observed 1 Data cables For the connection between instruments resp their interfaces and external devices computer printer etc su...

Page 5: ...x ambient temperature 0 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Color techno brown Cabinet W 285 H 125 D 380 mm Subject to change without notice 08 00 This microprocessor...

Page 6: ...f an extension cord without a protective conductor The mains line plug must be inserted before connec tions are made to measuring circuits The grounded accessible metal parts case sockets jacks and th...

Page 7: ...helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain intervals Only in this way is it largel...

Page 8: ...it a true signal display The display of sinusoidal signals within the bandwidth limits causes no prob lems but an increasing error in measurement due to gain reduction must be taken into account when...

Page 9: ...caus ing damage to the input of the oscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case...

Page 10: ...asurement of the rise or fall time is not limited to the trace dimensions shown in the above diagram It is only particularly simple in this way In principle it is possible to measure in any display po...

Page 11: ...Hz There fore the derating curve of the attenuator probe type con cerned must be taken into account The selection of the ground point on the test object is important when displaying small signal volta...

Page 12: ...iver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile memories These can be used by the ope...

Page 13: ...with this control knob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a h...

Page 14: ...aid of the X POS control As the X expansion results in a higher timebase speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over Please note...

Page 15: ...hannel II followed by XY Except the cursor lines which may be active all other readout information including the trigger point symbol are switched off In addition to all trigger and timebase related c...

Page 16: ...l switched off NR high frequency noise rejected LF low pass filter cuts off frequencies above approx 1 5kHz TVL TV signal line pulse triggering trigger point symbol switched off TVF TV signal frame pu...

Page 17: ...e TIME DIV 24 control knob is operative only for this timebase The readout then only displays the A time coefficient The timebase settings for this condition are stored if the timebase mode is changed...

Page 18: ...switch and B vernier VAR LED on In the latter case the TIME DIV knob can be used in the same way as de scribed before under A timebase condition Underneath the front panel sector described above the...

Page 19: ...r related con trols assumes that the readout is visible and the component tester is switched off 35 MENU Pushbutton Pressing and holding the pushbutton activates the display of the MAIN MENU It contai...

Page 20: ...as one signal sum or difference As the result can only be determined if both calibrated deflection coefficients are equal the CHI II 36 selection function is deactivated In that case the readout indic...

Page 21: ...ted to the oscilloscope input s Check that the device under test is switched off and connect the test lead s to the test point s Then switch on the instrument and afterwards the device under test The...

Page 22: ...rm distortion Prerequisite for this HF compensation is a square wave generator with fast risetime typically 4ns and low output impedance approx 50 providing 0 2V at a frequency of approx 1MHz The cali...

Page 23: ...this operation mode is the DUAL XY 18 pushbutton please note section Con trols and Readout In XY mode the timebase is deactivated The signal applied to the input of channel II front panel marking HOR...

Page 24: ...ng for one period in div In the example illustrated t 3div and T 10div The phase difference in degrees is calculated from Relatively small phase angles at lower frequencies can be measured more accura...

Page 25: ...ode for the entry into difficult measuring problems e g when the test signal is unknown relating to amplitude frequency or shape Presetting of all parameters is now possible with automatic triggering...

Page 26: ...5div height The polarity of the synchronization pulse is critical for the slope selection If the displayed sync pulses are above the picture field contents leading edge positive going then the slope...

Page 27: ...vated As the external trigger signal applied at the TRIG EXT socket normally has no relation to the signal height of the displayed signal the trigger point symbol is switched off The external trigger...

Page 28: ...nt This allows the sweep to begin on any portion of a signal except the first millimeters If the display shows jitter it is possible to select triggering of the B timebase after the elapsed delay time...

Page 29: ...lugs and with test prod alligator clip or sprung hook are required The test leads are connected as described in section Con trols and Readout Test Procedure Caution Do not test any component in live c...

Page 30: ...r a transistor the figures b e and b c are important The figure e c can vary but a vertical line only shows short circuit condition These transistor test patterns are valid in most cases but there are...

Page 31: ...devices In case of disregard of the safety warnings contained in this manual HAMEG refuses any liability regarding personal injury and or damage of equip ment Operation The oscilloscope is supplied w...

Page 32: ...Subject to change without notice 32 Front Panel HM1004 3...

Page 33: ...33 Subject to change without notice...

Page 34: ...Subject to change without notice 34...

Page 35: ...MANUAL HANDBUCH MANUEL...

Page 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: