Hameg HM 1004-3 Series Manual Download Page 12

Subject to change without notice

12

Controls and Readout

In DUAL mode the CURSOR lines are related to the signal
which is used for internal triggering.

If the signal height is insufficient, the CURSOR lines do
not change.

(3) RM 

-

The remote control mode can be switched on or off via
the RS232 interface. In the latter case the

 “RM” LED 

is

lit and the electronically selectable controls on front panel
are inactive. This state can be left by depressing the

AUTO SET 

pushbutton provided it was not deactivated

via the interface.

(4) INTENS - READOUT

Knob with associated pushbutton and LEDs.

This control knob is for adjusting the A and B traces and
readout intensity. Turning this knob clockwise increases
and turning it counterclockwise decreases the intensity.

The

 READOUT 

pushbutton below is for selecting the

function in two ways.

Depending on the actual timebase mode and with the
readout

 (RO) 

not switched off, briefly pressing the

READOUT 

pushbutton switches over the

 INTENS 

knob

function indicated by a LED in the sequences:

A - RO - A in condition A timebase,
A - RO - B - A if alternate timebase mode is present and
B - RO - B in condition B timebase.

XY mode: A - RO - A.
Component Test: A - RO - A.

Pressing and holding the

 READOUT 

pushbutton switches

the readout on or off. In readout off condition the

 INTENS

knob function can consequently not be set to

 RO

. Briefly

pressing the pushbutton causes the following sequences:

condition

sequence

A timebase

A - A

Alternate A/B

A - B - A

B timebase

B - B

XY mode

A - A

Component Test

A - A

Switching the readout off, may be required if interference
is visible on the signal(s). Such interference may also
originate from the chopper generator if the instrument is
operated in chopped

 DUAL 

mode.

With the exception of the letters “CT” all other READ-
OUT information is switched off in COMPONENT TEST
mode.
All INTENS settings are stored after the instrument is
switched off.

The

 AUTOSET 

function switches the readout on and

selects A timebase mode (A-LED lit). The

 INTENS 

setting

for each function is automatically set to the mean value,
if less intensity was previously selected.

(5) TR

The trace rotation control can be adjusted with a small
screwdriver (

please note “trace rotation TR”

)

(6) FOCUS

This control knob effects both the trace and the readout
sharpness.

(7) SAVE / RECALL

The instrument contains 9 non volatile memories. These
can be used by the operator to save instrument settings
and to recall them. This relates to all settings with the
exception of

 FOCUS, TR 

(trace rotation) and the calibrator

frequency pushbutton.

Press the

 SAVE 

pushbutton briefly to start the save

procedure. The readout then indicates the letter

  “S”

followed by a cipher between 1 and 9, indicating the
memory location. If the instrument settings stored in this
memory location must not be overwritten, briefly press
the

 SAVE 

or the

 RECALL 

pushbutton to select another

memory location. Each time the

 SAVE 

pushbutton is

briefly pressed the memory location cipher increases
until the location number 9 is reached. The

 RECALL

pushbutton function is similar but decreases the memory
location cipher until 1 is reached. Press and hold

 SAVE 

for

approx. 3 seconds to write the instruments settings in the
memory and to switch the associated readout informa-
tion (e.g. “S8”) off.

To recall a front panel setup, start that procedure by
briefly pressing the

 RECALL 

pushbutton. The readout

then indicates the letter

  “R” 

and the memory location

number. If required, select a different memory location as
described above. Recall the settings by pressing and
holding the RECALL pushbutton for approx. 3 seconds.

Attention:
Make sure that the signal to be displayed is similar to
the one that was present when the settings were
stored. If the signal is different (frequency, amplitude)
to the one during storage then a distorted display may
result.

If the

 SAVE 

or the

 RECALL 

pushbutton was depressed

inadvertently, briefly press both pushbuttons at the same
time or wait approx. 10 seconds without pressing either
pushbutton to exit that function.

Switching the instrument off automatically stores the
actual settings in memory location 9, with the effect that
different settings previously stored in this location get
lost. To prevent this, RECALL 9 before switching the
instrument off.

Attention!
Both pushbuttons have a second function if the
instrument is switched to menu operation. Please
note "MENU".

The setting controls and LED’s for the Y amplifiers,
modes, triggering and timebases are located under-
neath the sector of the front panel described before.

Summary of Contents for HM 1004-3 Series

Page 1: ...Instruments Oscilloscope HM 1004 3 01 02 03 ENGLISH MANUAL HANDBUCH MANUEL...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...res 23 Phase difference measurement in DUAL mode Yt 24 Phase difference measurement in DUAL mode 24 Measurement of amplitude modulation 24 Triggering and timebase 25 Automatic Peak value Triggering 25...

Page 4: ...ollowing hints and conditions regarding emission and immunity should be observed 1 Data cables For the connection between instruments resp their interfaces and external devices computer printer etc su...

Page 5: ...x ambient temperature 0 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Color techno brown Cabinet W 285 H 125 D 380 mm Subject to change without notice 08 00 This microprocessor...

Page 6: ...f an extension cord without a protective conductor The mains line plug must be inserted before connec tions are made to measuring circuits The grounded accessible metal parts case sockets jacks and th...

Page 7: ...helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain intervals Only in this way is it largel...

Page 8: ...it a true signal display The display of sinusoidal signals within the bandwidth limits causes no prob lems but an increasing error in measurement due to gain reduction must be taken into account when...

Page 9: ...caus ing damage to the input of the oscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case...

Page 10: ...asurement of the rise or fall time is not limited to the trace dimensions shown in the above diagram It is only particularly simple in this way In principle it is possible to measure in any display po...

Page 11: ...Hz There fore the derating curve of the attenuator probe type con cerned must be taken into account The selection of the ground point on the test object is important when displaying small signal volta...

Page 12: ...iver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile memories These can be used by the ope...

Page 13: ...with this control knob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a h...

Page 14: ...aid of the X POS control As the X expansion results in a higher timebase speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over Please note...

Page 15: ...hannel II followed by XY Except the cursor lines which may be active all other readout information including the trigger point symbol are switched off In addition to all trigger and timebase related c...

Page 16: ...l switched off NR high frequency noise rejected LF low pass filter cuts off frequencies above approx 1 5kHz TVL TV signal line pulse triggering trigger point symbol switched off TVF TV signal frame pu...

Page 17: ...e TIME DIV 24 control knob is operative only for this timebase The readout then only displays the A time coefficient The timebase settings for this condition are stored if the timebase mode is changed...

Page 18: ...switch and B vernier VAR LED on In the latter case the TIME DIV knob can be used in the same way as de scribed before under A timebase condition Underneath the front panel sector described above the...

Page 19: ...r related con trols assumes that the readout is visible and the component tester is switched off 35 MENU Pushbutton Pressing and holding the pushbutton activates the display of the MAIN MENU It contai...

Page 20: ...as one signal sum or difference As the result can only be determined if both calibrated deflection coefficients are equal the CHI II 36 selection function is deactivated In that case the readout indic...

Page 21: ...ted to the oscilloscope input s Check that the device under test is switched off and connect the test lead s to the test point s Then switch on the instrument and afterwards the device under test The...

Page 22: ...rm distortion Prerequisite for this HF compensation is a square wave generator with fast risetime typically 4ns and low output impedance approx 50 providing 0 2V at a frequency of approx 1MHz The cali...

Page 23: ...this operation mode is the DUAL XY 18 pushbutton please note section Con trols and Readout In XY mode the timebase is deactivated The signal applied to the input of channel II front panel marking HOR...

Page 24: ...ng for one period in div In the example illustrated t 3div and T 10div The phase difference in degrees is calculated from Relatively small phase angles at lower frequencies can be measured more accura...

Page 25: ...ode for the entry into difficult measuring problems e g when the test signal is unknown relating to amplitude frequency or shape Presetting of all parameters is now possible with automatic triggering...

Page 26: ...5div height The polarity of the synchronization pulse is critical for the slope selection If the displayed sync pulses are above the picture field contents leading edge positive going then the slope...

Page 27: ...vated As the external trigger signal applied at the TRIG EXT socket normally has no relation to the signal height of the displayed signal the trigger point symbol is switched off The external trigger...

Page 28: ...nt This allows the sweep to begin on any portion of a signal except the first millimeters If the display shows jitter it is possible to select triggering of the B timebase after the elapsed delay time...

Page 29: ...lugs and with test prod alligator clip or sprung hook are required The test leads are connected as described in section Con trols and Readout Test Procedure Caution Do not test any component in live c...

Page 30: ...r a transistor the figures b e and b c are important The figure e c can vary but a vertical line only shows short circuit condition These transistor test patterns are valid in most cases but there are...

Page 31: ...devices In case of disregard of the safety warnings contained in this manual HAMEG refuses any liability regarding personal injury and or damage of equip ment Operation The oscilloscope is supplied w...

Page 32: ...Subject to change without notice 32 Front Panel HM1004 3...

Page 33: ...33 Subject to change without notice...

Page 34: ...Subject to change without notice 34...

Page 35: ...MANUAL HANDBUCH MANUEL...

Page 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: