Hameg HM 1004-3 Series Manual Download Page 10

Subject to change without notice

10

Type of signal voltage

set time coefficient Tc = 0.2s/div,
required wavelength L = 1:0.2 = 5div.
Displayed ripple wavelength L = 1div,
set time coefficient Tc = 10ms/div,
required ripple freq. F = 1:(1x10x10-3) = 100Hz.
TV-Line frequency F = 15625Hz,
set time coefficient Tc = 10µs/div,
required wavelength L = 1:(15 625x10-5) = 6.4div.
Sine wavelength L = min. 4div, max. 10div,
Frequency F = 1kHz,
max. time coefficient Tc = 1:(4x103) = 0.25ms/div,
min. time coefficient Tc = 1:(10x103) = 0.1ms/div,
set time coefficient Tc = 0.2ms/div,
required wavelength L = 1:(103x0.2x10-3) = 5div.
Displayed wavelength L = 0.8div,
set time coefficient Tc = 0.5µs/div,
pressed X-MAG. (x10) button: Tc = 0.05µs/div,
required rec. freq. F = 1:(0.8x0.05x10-6) = 25MHz,
required period T = 1:(25x106) = 40ns.

If the time is relatively short as compared with the complete
signal period, an expanded time scale should always be applied
(X-MAG. (x10) active). In this case, the time interval of interest
can be shifted to the screen center using the X-POS. control.

When investigating pulse or square waveforms, the critical
feature is the risetime of the voltage step. To ensure that
transients, ramp-offs, and bandwidth limits do not unduly
influence the measuring accuracy, the risetime is generally
measured between 10% and 90% of the vertical pulse height.
For measurement, adjust the Y deflection coefficient using its
variable function (uncalibrated) together with the Y-POS.
control so that the pulse height is precisely aligned with the
0% and 100% lines of the internal graticule. The 10% and
90% points of the signal will now coincide with the 10% and
90% graticule lines. The risetime is given by the product of the
horizontal distance in div between these two coincident
points and the calibrated time coefficient setting. The fall time
of a pulse can also be measured by using this method.

The following figure shows correct positioning of the oscillo-
scope trace for accurate risetime measurement.

With a time coefficient of 5ns/div (X x10 magnification active),
the example shown in the above figure results in a total
measured risetime of

ttot = 1.6div x 5ns/div = 

8ns

When very fast risetimes are being measured, the risetimes
of the oscilloscope amplifier and of the attenuator probe has
to be deducted from the measured time value. The risetime
of the signal can be calculated using the following formula.

t

r

 = 

 t

tot2

 - t

osc2

 - t

p2

In this ttot is the total measured risetime, tosc is the risetime

of the oscilloscope amplifier (HM1004-3 approx. 3.5ns) and tp

the risetime of the probe (e.g. = 2ns). If ttot is greater than

34ns, then ttot can be taken as the risetime of the pulse, and

calculation is unnecessary.

Calculation of the example in the figure above results in a
signal risetime

tr = 

 8

2

 - 3,5

2

 - 2

6,9ns

The measurement of the rise or fall time is not limited to the
trace dimensions shown in the above diagram. It is only
particularly simple in this way. In principle it is possible to
measure in any display position and at any signal amplitude.
It is only important that the full height of the signal edge of
interest is visible in its full length at not too great steepness
and that the horizontal distance at 10% and 90% of the
amplitude is measured. If the edge shows rounding or over-
shooting, the 100% should not be related to the peak values
but to the mean pulse heights. Breaks or peaks (glitches) next
to the edge are also not taken into account. With very severe
transient distortions, the rise and fall time measurement has
little meaning. For amplifiers with approximately constant
group delay (therefore good pulse transmission performance)
the following numerical relationship between rise time

 tr (in

ns)

 and bandwidth 

B (in MHz)

 applies:

Connection of Test Signal

In most cases briefly depressing the

 AUTO SET 

causes a

useful signal related instrument setting. The following expla-
nations refer to special applications and/or signals, demand-
ing a manual instrument setting. 

The description of the

controls is explained in the section “controls and readout”

.

Caution:
When connecting unknown signals to the oscilloscope
input, always use automatic triggering and set the
input coupling switch to AC (readout). The attenuator
should initially be set to 20V/div.

Sometimes the trace will disappear after an input signal has
been applied. Then a higher deflection coefficient (lower input
sensitivity) must be chosen until the vertical signal height is
only 3-8div. With a signal amplitude greater than 160Vpp and
the deflection coefficient 

(VOLTS/DIV.)

 in calibrated condi-

tion, an attenuator probe must be inserted before the vertical
input. If, after applying the signal, the trace is nearly blanked,
the period of the signal is probably substantially longer than
the set time deflection coefficient (

TIME/DIV.

). It should be

switched to an adequately larger time coefficient.

The signal to be displayed can be connected directly to the Y-
input of the oscilloscope with a shielded test cable such as

HZ32 

or

 HZ34

, or reduced through a x10 or x100 attenuator

probe. The use of test cables with high impedance circuits is
only recommended for relatively low frequencies (up to
approx. 50kHz). For higher frequencies, the signal source
must be of low impedance, i.e. matched to the characteristic
resistance of the cable (as a rule 50

). Especially when

transmitting square and pulse signals, a resistor equal to the
characteristic impedance of the cable must also be connected
across the cable directly at the Y-input of the oscilloscope.
When using a 50

 cable such as the

 HZ34

, a 50

 through

termination type

 HZ22

 is available from

 HAMEG

. When

transmitting square signals with short rise times, transient
phenomena on the edges and top of the signal may become

Summary of Contents for HM 1004-3 Series

Page 1: ...Instruments Oscilloscope HM 1004 3 01 02 03 ENGLISH MANUAL HANDBUCH MANUEL...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...res 23 Phase difference measurement in DUAL mode Yt 24 Phase difference measurement in DUAL mode 24 Measurement of amplitude modulation 24 Triggering and timebase 25 Automatic Peak value Triggering 25...

Page 4: ...ollowing hints and conditions regarding emission and immunity should be observed 1 Data cables For the connection between instruments resp their interfaces and external devices computer printer etc su...

Page 5: ...x ambient temperature 0 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Color techno brown Cabinet W 285 H 125 D 380 mm Subject to change without notice 08 00 This microprocessor...

Page 6: ...f an extension cord without a protective conductor The mains line plug must be inserted before connec tions are made to measuring circuits The grounded accessible metal parts case sockets jacks and th...

Page 7: ...helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain intervals Only in this way is it largel...

Page 8: ...it a true signal display The display of sinusoidal signals within the bandwidth limits causes no prob lems but an increasing error in measurement due to gain reduction must be taken into account when...

Page 9: ...caus ing damage to the input of the oscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case...

Page 10: ...asurement of the rise or fall time is not limited to the trace dimensions shown in the above diagram It is only particularly simple in this way In principle it is possible to measure in any display po...

Page 11: ...Hz There fore the derating curve of the attenuator probe type con cerned must be taken into account The selection of the ground point on the test object is important when displaying small signal volta...

Page 12: ...iver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile memories These can be used by the ope...

Page 13: ...with this control knob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a h...

Page 14: ...aid of the X POS control As the X expansion results in a higher timebase speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over Please note...

Page 15: ...hannel II followed by XY Except the cursor lines which may be active all other readout information including the trigger point symbol are switched off In addition to all trigger and timebase related c...

Page 16: ...l switched off NR high frequency noise rejected LF low pass filter cuts off frequencies above approx 1 5kHz TVL TV signal line pulse triggering trigger point symbol switched off TVF TV signal frame pu...

Page 17: ...e TIME DIV 24 control knob is operative only for this timebase The readout then only displays the A time coefficient The timebase settings for this condition are stored if the timebase mode is changed...

Page 18: ...switch and B vernier VAR LED on In the latter case the TIME DIV knob can be used in the same way as de scribed before under A timebase condition Underneath the front panel sector described above the...

Page 19: ...r related con trols assumes that the readout is visible and the component tester is switched off 35 MENU Pushbutton Pressing and holding the pushbutton activates the display of the MAIN MENU It contai...

Page 20: ...as one signal sum or difference As the result can only be determined if both calibrated deflection coefficients are equal the CHI II 36 selection function is deactivated In that case the readout indic...

Page 21: ...ted to the oscilloscope input s Check that the device under test is switched off and connect the test lead s to the test point s Then switch on the instrument and afterwards the device under test The...

Page 22: ...rm distortion Prerequisite for this HF compensation is a square wave generator with fast risetime typically 4ns and low output impedance approx 50 providing 0 2V at a frequency of approx 1MHz The cali...

Page 23: ...this operation mode is the DUAL XY 18 pushbutton please note section Con trols and Readout In XY mode the timebase is deactivated The signal applied to the input of channel II front panel marking HOR...

Page 24: ...ng for one period in div In the example illustrated t 3div and T 10div The phase difference in degrees is calculated from Relatively small phase angles at lower frequencies can be measured more accura...

Page 25: ...ode for the entry into difficult measuring problems e g when the test signal is unknown relating to amplitude frequency or shape Presetting of all parameters is now possible with automatic triggering...

Page 26: ...5div height The polarity of the synchronization pulse is critical for the slope selection If the displayed sync pulses are above the picture field contents leading edge positive going then the slope...

Page 27: ...vated As the external trigger signal applied at the TRIG EXT socket normally has no relation to the signal height of the displayed signal the trigger point symbol is switched off The external trigger...

Page 28: ...nt This allows the sweep to begin on any portion of a signal except the first millimeters If the display shows jitter it is possible to select triggering of the B timebase after the elapsed delay time...

Page 29: ...lugs and with test prod alligator clip or sprung hook are required The test leads are connected as described in section Con trols and Readout Test Procedure Caution Do not test any component in live c...

Page 30: ...r a transistor the figures b e and b c are important The figure e c can vary but a vertical line only shows short circuit condition These transistor test patterns are valid in most cases but there are...

Page 31: ...devices In case of disregard of the safety warnings contained in this manual HAMEG refuses any liability regarding personal injury and or damage of equip ment Operation The oscilloscope is supplied w...

Page 32: ...Subject to change without notice 32 Front Panel HM1004 3...

Page 33: ...33 Subject to change without notice...

Page 34: ...Subject to change without notice 34...

Page 35: ...MANUAL HANDBUCH MANUEL...

Page 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: