Hameg HM 1004-3 Series Manual Download Page 24

Subject to change without notice

24

First Time Operation

Should both input voltages be missing or fail in the X-Y mode,
a very bright dot is displayed on the screen. This dot can burn
into the phosphor at a too high brightness setting (

INTENS.

knob

) which causes either a lasting loss of brightness, or in

the extreme case, complete destruction of the phosphor at
this point.

Phase difference measurement
in DUAL mode (Yt)

Phase differences between two input signals of the same
frequency and shape can be measured very simply on the
screen in Dual mode. The timebase should be triggered by the
reference signal (phase position 0).  The other signal can then
have a leading or lagging phase angle.

For greatest accuracy adjust the timebase for slightly over
one period and approximately the same height of both signals
on the screen. The Y deflection coefficients, the timebase
coefficients and the trigger level setting can be used for this
adjustment, without influence on the result. Both base lines
are set onto the horizontal graticule center line using the Y-
POS. knobs before the measurement. With sinusoidal sig-
nals, observe the zero (crossover point) transitions; the sine
peaks are less accurate.

If a sine signal is noticeably distorted by even harmonics, or
if a DC voltage is present, AC coupling is recommended for
both channels. If it is a question of pulses of the same shape,
read off at steep edges.

It must be noted that the phase difference cannot be deter-
mined if alternate triggering is selected.

Phase difference measurement in DUAL mode

t = horizontal spacing of the zero transitions in div
T = horizontal spacing for one period in div

In the example illustrated, t = 3div and T = 10div.

The phase difference in degrees is calculated from

Relatively small phase angles at lower frequencies can be
measured more accurately in the X-Y mode with Lissajous
figures.

Measurement of amplitude modulation

The momentary amplitude u at time t of a HF-carrier voltage,
which is amplitude modulated without distortion by a sinusoidal
AF voltage, is in accordance with the equation

where

U

T

= unmodulated carrier amplitude

= 2

π

F = angular carrier frequency

ω

= 2

π

f = modulation angular frequency

m

= modulation factor

The lower side frequency F-f  and the upper side frequency
F+f  arise because of the modulation apart from the carrier
frequency F.

Figure 1
Amplitude and frequency spectrum for AM display (m = 50%)

The display of an amplitude-modulated HF oscillation can be
evaluated with the oscilloscope provided the frequency spec-
trum is inside the oscilloscope bandwidth. The timebase is set
so that several cycles of the modulation frequency are visible.
Strictly speaking, triggering should be external with modula-
tion frequency (from the AF generator or a demodulator).
However, internal triggering is frequently possible with nor-
mal triggering using a suitable trigger level setting and possi-
bly also using the time vernier (variable) adjustment.

Oscilloscope setting for a signal according to figure 2:

Y: CH. I; 20mV/div; AC.

TIME/DIV.: 0.2ms/div.

Triggering: Normal; with LEVEL-setting; internal
(or external) triggering.

Figure 2
Amplitude modulated oscillation:
F = 1 MHz; f = 1 kHz;
m = 50 %; UT = 28.3 mVrms.

If the two values a and b are read from the screen, the
modulation factor is calculated from

where

a = UT (1+m) 

and

 b = UT (1-m)

.

The variable controls for amplitude and time can be set
arbitrarily in the modulation factor measurement. Their posi-
tion does not influence the result.

Summary of Contents for HM 1004-3 Series

Page 1: ...Instruments Oscilloscope HM 1004 3 01 02 03 ENGLISH MANUAL HANDBUCH MANUEL...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...res 23 Phase difference measurement in DUAL mode Yt 24 Phase difference measurement in DUAL mode 24 Measurement of amplitude modulation 24 Triggering and timebase 25 Automatic Peak value Triggering 25...

Page 4: ...ollowing hints and conditions regarding emission and immunity should be observed 1 Data cables For the connection between instruments resp their interfaces and external devices computer printer etc su...

Page 5: ...x ambient temperature 0 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Color techno brown Cabinet W 285 H 125 D 380 mm Subject to change without notice 08 00 This microprocessor...

Page 6: ...f an extension cord without a protective conductor The mains line plug must be inserted before connec tions are made to measuring circuits The grounded accessible metal parts case sockets jacks and th...

Page 7: ...helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain intervals Only in this way is it largel...

Page 8: ...it a true signal display The display of sinusoidal signals within the bandwidth limits causes no prob lems but an increasing error in measurement due to gain reduction must be taken into account when...

Page 9: ...caus ing damage to the input of the oscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case...

Page 10: ...asurement of the rise or fall time is not limited to the trace dimensions shown in the above diagram It is only particularly simple in this way In principle it is possible to measure in any display po...

Page 11: ...Hz There fore the derating curve of the attenuator probe type con cerned must be taken into account The selection of the ground point on the test object is important when displaying small signal volta...

Page 12: ...iver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile memories These can be used by the ope...

Page 13: ...with this control knob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a h...

Page 14: ...aid of the X POS control As the X expansion results in a higher timebase speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over Please note...

Page 15: ...hannel II followed by XY Except the cursor lines which may be active all other readout information including the trigger point symbol are switched off In addition to all trigger and timebase related c...

Page 16: ...l switched off NR high frequency noise rejected LF low pass filter cuts off frequencies above approx 1 5kHz TVL TV signal line pulse triggering trigger point symbol switched off TVF TV signal frame pu...

Page 17: ...e TIME DIV 24 control knob is operative only for this timebase The readout then only displays the A time coefficient The timebase settings for this condition are stored if the timebase mode is changed...

Page 18: ...switch and B vernier VAR LED on In the latter case the TIME DIV knob can be used in the same way as de scribed before under A timebase condition Underneath the front panel sector described above the...

Page 19: ...r related con trols assumes that the readout is visible and the component tester is switched off 35 MENU Pushbutton Pressing and holding the pushbutton activates the display of the MAIN MENU It contai...

Page 20: ...as one signal sum or difference As the result can only be determined if both calibrated deflection coefficients are equal the CHI II 36 selection function is deactivated In that case the readout indic...

Page 21: ...ted to the oscilloscope input s Check that the device under test is switched off and connect the test lead s to the test point s Then switch on the instrument and afterwards the device under test The...

Page 22: ...rm distortion Prerequisite for this HF compensation is a square wave generator with fast risetime typically 4ns and low output impedance approx 50 providing 0 2V at a frequency of approx 1MHz The cali...

Page 23: ...this operation mode is the DUAL XY 18 pushbutton please note section Con trols and Readout In XY mode the timebase is deactivated The signal applied to the input of channel II front panel marking HOR...

Page 24: ...ng for one period in div In the example illustrated t 3div and T 10div The phase difference in degrees is calculated from Relatively small phase angles at lower frequencies can be measured more accura...

Page 25: ...ode for the entry into difficult measuring problems e g when the test signal is unknown relating to amplitude frequency or shape Presetting of all parameters is now possible with automatic triggering...

Page 26: ...5div height The polarity of the synchronization pulse is critical for the slope selection If the displayed sync pulses are above the picture field contents leading edge positive going then the slope...

Page 27: ...vated As the external trigger signal applied at the TRIG EXT socket normally has no relation to the signal height of the displayed signal the trigger point symbol is switched off The external trigger...

Page 28: ...nt This allows the sweep to begin on any portion of a signal except the first millimeters If the display shows jitter it is possible to select triggering of the B timebase after the elapsed delay time...

Page 29: ...lugs and with test prod alligator clip or sprung hook are required The test leads are connected as described in section Con trols and Readout Test Procedure Caution Do not test any component in live c...

Page 30: ...r a transistor the figures b e and b c are important The figure e c can vary but a vertical line only shows short circuit condition These transistor test patterns are valid in most cases but there are...

Page 31: ...devices In case of disregard of the safety warnings contained in this manual HAMEG refuses any liability regarding personal injury and or damage of equip ment Operation The oscilloscope is supplied w...

Page 32: ...Subject to change without notice 32 Front Panel HM1004 3...

Page 33: ...33 Subject to change without notice...

Page 34: ...Subject to change without notice 34...

Page 35: ...MANUAL HANDBUCH MANUEL...

Page 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: