Hameg HM 1004-3 Series Manual Download Page 28

Subject to change without notice

28

Triggering and Timebase

Auto Set

calibration detent (fully ccw), otherwise the brightness of the
display is reduced drastically. The function is shown in the
following figures.

Fig. 1 shows a case where the holdoff control is in the minimum

position and various different waveforms are overlapped on
the screen, making the signal observation unsuccessful.

Fig. 2 shows a case where only the desired parts of the signal are

stably displayed.

B-Timebase (2nd Timebase)/
Triggering after Delay

Please note instrument specific information in section
“Controls and readout”. The most important controls
regarding this mode are 

A/ALT - B (25), DEL.TRIG. - VAR.

(26), TIME/DIV. (24), DEL. POS.- HO (23)

 and 

 and 

 and 

 and 

 and 

TRS (9).

As mentioned before, triggering starts the timebase sweep
and unblanks the beam. After the maximum X deflection to
the right, the beam is blanked and flies back to the (left) start
position. After the hold off time has elapsed the sweep is
started automatically by the automatic trigger or the next
trigger signal. In normal triggering mode the automatic trigger
is switched off and the sweep will only start on receipt of a
trigger signal. As the trigger point is always at the trace start
position, trace expansion in X direction with the aid of the
timebase is limited to the display always from the trace start.

Parts of the signal to be expanded which are displayed near
the trace end (right side of the screen) are lost when the
timebase speed is increased (time coefficient reduced). The
X expansion by 10 may also be used, but again the expansion
factor is limited. The B timebase solves such problems.

First the A time coefficient should be in a suitable setting to
display a complete signal period. After setting the instrument
to alternate timebase mode (A and B), the signal part to be
expanded must be selected with the aid of the intensified
sector on the A timebase trace. The B timebase deflection
coefficient should be at least a tenth of the A time coefficient.

Higher A/B ratios (e.g. A: 1ms/div, B: 1µs/div = 1000fold
expansion) result in a drastically reduced B trace intensity.
Within certain limits this can be compensated by the intensity
setting. To avoid high contrast between the A and B timebase
trace, change from alternate (A and B) to B timebase mode.
Then the A trace is invisible.

The start position of the intensified sector on the A trace can
be stated as the delay time. It is related to the A trace start and
the A time coefficient.

The delay function delays the B trace start by a variable time
from the A trigger point. This allows the sweep to begin on any
portion of a signal (except the first millimeters). If the display
shows jitter, it is possible to select triggering of the B
timebase after the elapsed delay time.

Auto Set

The instrument specific information regarding this function is
part of the section

  “Controls and Readout” paragraph

AUTO SET (2)

. As also mentioned in that section all controls

are electronically selected with the exception of the

 POWER

and calibrator frequency (CAL 1kHz/1MHz) pushbuttons as
well as the

 FOCUS 

and

 TR 

(trace rotation) control.

Thus automatic signal related instrument set up in Yt (timebase)
mode is possible. In most cases no additional manual instru-
ment setting is required.

Briefly pressing the

 AUTO SET 

pushbutton causes the instru-

ment to switch over to the last Yt mode settings regarding CH
I, CH II and DUAL. If the instrument was operated in Yt mode,
the actual setting will not be affected with the exception of
ADD mode which will be switched off. At the same time the
attenuator(s) (VOLTS/DIV) are automatically set for a signal
display height of  approx. 6 div in mono channel mode or if in
DUAL mode for approx. 4 div height for each channel. This and
the following explanation regarding the automatic time coef-
ficient setting assumes that the pulse duty factor of the input
signal is approx. 1:1. The time deflection coefficient is also set
automatically for a display of approx. 2 signal periods. The
timebase setting occurs randomly if complex signals consist-
ing several frequencies e.g. video signals are present.

If cursor voltage measurement is selected, AUTOSET also
affects the position of the CURSOR lines. Please note
AUTOSET (2) in section ”Controls and Readout”.

AUTO SET 

sets the instrument automatically to the following

operating conditions:

• AC or DC input coupling unaltered

or in GD condition the previously used setting

• Internal triggering (channel I or channel II)
• Automatic peak (value) triggering
• Trigger level in electrical midrange position
• Y deflection coefficient(s) calibrated
• A timebase deflection coefficient calibrated
• AC trigger coupling (except DC trigger coupling)
• Alternate or B timebase mode switched off
• X x10 magnifier switched off
• Automatic X and Y position settings
• A visible trace and readout intensities.

If DC trigger coupling had been selected, AC trigger coupling
will not be chosen and the automatic trigger is operative
without the peak value detection.

The X position is set to the CRT center as well as the Y position
in CH I or CH II mode.

 Small deviations due to tolerances

of the CRT and other components are unavoidable. 

In

DUAL mode the channel I trace is set to the upper half and the
channel II trace to the lower half of the CRT.

The 1mV/div and 2mV/div deflection coefficients will not be

Summary of Contents for HM 1004-3 Series

Page 1: ...Instruments Oscilloscope HM 1004 3 01 02 03 ENGLISH MANUAL HANDBUCH MANUEL...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...res 23 Phase difference measurement in DUAL mode Yt 24 Phase difference measurement in DUAL mode 24 Measurement of amplitude modulation 24 Triggering and timebase 25 Automatic Peak value Triggering 25...

Page 4: ...ollowing hints and conditions regarding emission and immunity should be observed 1 Data cables For the connection between instruments resp their interfaces and external devices computer printer etc su...

Page 5: ...x ambient temperature 0 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Color techno brown Cabinet W 285 H 125 D 380 mm Subject to change without notice 08 00 This microprocessor...

Page 6: ...f an extension cord without a protective conductor The mains line plug must be inserted before connec tions are made to measuring circuits The grounded accessible metal parts case sockets jacks and th...

Page 7: ...helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain intervals Only in this way is it largel...

Page 8: ...it a true signal display The display of sinusoidal signals within the bandwidth limits causes no prob lems but an increasing error in measurement due to gain reduction must be taken into account when...

Page 9: ...caus ing damage to the input of the oscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case...

Page 10: ...asurement of the rise or fall time is not limited to the trace dimensions shown in the above diagram It is only particularly simple in this way In principle it is possible to measure in any display po...

Page 11: ...Hz There fore the derating curve of the attenuator probe type con cerned must be taken into account The selection of the ground point on the test object is important when displaying small signal volta...

Page 12: ...iver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile memories These can be used by the ope...

Page 13: ...with this control knob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a h...

Page 14: ...aid of the X POS control As the X expansion results in a higher timebase speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over Please note...

Page 15: ...hannel II followed by XY Except the cursor lines which may be active all other readout information including the trigger point symbol are switched off In addition to all trigger and timebase related c...

Page 16: ...l switched off NR high frequency noise rejected LF low pass filter cuts off frequencies above approx 1 5kHz TVL TV signal line pulse triggering trigger point symbol switched off TVF TV signal frame pu...

Page 17: ...e TIME DIV 24 control knob is operative only for this timebase The readout then only displays the A time coefficient The timebase settings for this condition are stored if the timebase mode is changed...

Page 18: ...switch and B vernier VAR LED on In the latter case the TIME DIV knob can be used in the same way as de scribed before under A timebase condition Underneath the front panel sector described above the...

Page 19: ...r related con trols assumes that the readout is visible and the component tester is switched off 35 MENU Pushbutton Pressing and holding the pushbutton activates the display of the MAIN MENU It contai...

Page 20: ...as one signal sum or difference As the result can only be determined if both calibrated deflection coefficients are equal the CHI II 36 selection function is deactivated In that case the readout indic...

Page 21: ...ted to the oscilloscope input s Check that the device under test is switched off and connect the test lead s to the test point s Then switch on the instrument and afterwards the device under test The...

Page 22: ...rm distortion Prerequisite for this HF compensation is a square wave generator with fast risetime typically 4ns and low output impedance approx 50 providing 0 2V at a frequency of approx 1MHz The cali...

Page 23: ...this operation mode is the DUAL XY 18 pushbutton please note section Con trols and Readout In XY mode the timebase is deactivated The signal applied to the input of channel II front panel marking HOR...

Page 24: ...ng for one period in div In the example illustrated t 3div and T 10div The phase difference in degrees is calculated from Relatively small phase angles at lower frequencies can be measured more accura...

Page 25: ...ode for the entry into difficult measuring problems e g when the test signal is unknown relating to amplitude frequency or shape Presetting of all parameters is now possible with automatic triggering...

Page 26: ...5div height The polarity of the synchronization pulse is critical for the slope selection If the displayed sync pulses are above the picture field contents leading edge positive going then the slope...

Page 27: ...vated As the external trigger signal applied at the TRIG EXT socket normally has no relation to the signal height of the displayed signal the trigger point symbol is switched off The external trigger...

Page 28: ...nt This allows the sweep to begin on any portion of a signal except the first millimeters If the display shows jitter it is possible to select triggering of the B timebase after the elapsed delay time...

Page 29: ...lugs and with test prod alligator clip or sprung hook are required The test leads are connected as described in section Con trols and Readout Test Procedure Caution Do not test any component in live c...

Page 30: ...r a transistor the figures b e and b c are important The figure e c can vary but a vertical line only shows short circuit condition These transistor test patterns are valid in most cases but there are...

Page 31: ...devices In case of disregard of the safety warnings contained in this manual HAMEG refuses any liability regarding personal injury and or damage of equip ment Operation The oscilloscope is supplied w...

Page 32: ...Subject to change without notice 32 Front Panel HM1004 3...

Page 33: ...33 Subject to change without notice...

Page 34: ...Subject to change without notice 34...

Page 35: ...MANUAL HANDBUCH MANUEL...

Page 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: