Hameg HM 1004-3 Series Manual Download Page 13

13

Subject to change without notice

Controls and Readout

(8) TRS

The instrument contains a trace separation function which
is required in the alternate timebase mode to separate
the B timebase trace from the A timebase in Y direction.
Consequently this function is only available in alternate
timebase mode. After the TRS pushbutton was pressed
once the LED related to that pushbutton is lit.

The

 Y-POS. I 

control knob is then operative as vertical

position control for the trace of the B timebase. The
maximum position shift is approx. +/- 4 div. Without a
change of the

 Y-POS. I 

control the trace separation

function is switched off automatically after approx. 10
seconds. The trace separation function can also be left by
pressing the TRS pushbutton.

(9) Y-POS. I 

- Control knob with a double function.

Y-Position channel I:

The vertical trace position of channel I can be set with this
control knob. In ADD (addition) mode both (Y-POS. I and
Y-POS. II) control knobs are active.

Y-Position B-trace in alternate timebase mode:

In alternate timebase mode, this control knob can be
used to separate the B timebase trace from the A
timebase trace.

 Please note TRS (8)

.

DC voltage measurement:

If no signal is applied at the INPUT CHI (27), the vertical
trace position represents 0 Volt. This is the case if INPUT
CHI (27) or in addition (ADD) mode, both INPUT CHI (27)
and INPUT CHII (31), are set to GD (ground) and auto-
matic triggering (AT (11)) is present to make the trace
visible. The trace then can be set to vertical position
which is suited for the following DC voltage measure-
ment.
After switching GD (ground) off and selecting DC input
coupling, a DC signal applied at the input changes the
trace position in vertical direction. The DC voltage then
can be determined by taking the deflection coefficient,
the probe factor and the trace position change in respect
to the previous 0 Volt position into account.

”0 Volt” Symbol:

The determination of the ”0 Volt” position is not neces-
sary if the readout is switched on and the software
setting  ”DC Ref. = ON” is selected in the ”SETUP”
submenu ”Miscellaneous”. Then the ”

⊥⊥

⊥⊥

” symbol to the

left of the screen‘s vertical center line always indicates
the ”0 Volt” trace position in CHI and DUAL mode.

The ”0 Volt” position symbol (

⊥⊥

⊥⊥

) will not be displayed in

XY and ADD (addition) mode.

(10)Y-POS. II 

- Control knob.

The vertical trace position of channel II can be set with
this control knob. In ADD (addition) mode both (Y-POS. I
and Y-POS. II) control knobs are active. If the instrument
is set to XY mode this control knob is inactive and the X-
POS. knob must be used for a horizontal position shift.

DC voltage measurement:

If no signal is applied at the INPUT CHII (31), the vertical
trace position represents 0 Volt. This is the case if INPUT
CHII (31) or in addition (ADD) mode, both INPUT CHI (27)
and INPUT CHII (31), are set to GD (ground) and auto-
matic triggering (AT (11)) is present to make the trace
visible. The trace then can be set to vertical position
which is suited for the following DC voltage measure-
ment.

After switching GD (ground) off and selecting DC input
coupling, a DC signal applied at the input changes the
trace position in vertical direction. The DC voltage then
can be determined by taking the deflection coefficient,
the probe factor and the trace position change in respect
to the previous 0 Volt position into account.

”0 Volt” Symbol:

The determination of the ”0 Volt” position is not neces-
sary if the readout is switched on and the software
setting  ”DC Ref. = ON” is selected in the ”SETUP”
submenu ”Miscellaneous”. Then the ”

⊥⊥

⊥⊥

” symbol to the

left of the screen‘s vertical center line always indicates
the ”0 Volt” trace position in CHI and DUAL mode.

The ”0 Volt” position symbol (

⊥⊥

⊥⊥

) will not be displayed in

XY and ADD (addition) mode.

(11)NM - AT - 

(SLOPE)

Pushbutton with a double function and associated NM-
LED.

NM - AT selection:

Press and hold the pushbutton to switch over from
automatic (peak value) to normal triggering (NM LED
above the pushbutton lit) and vice versa. If the LED is
dark, automatic (peak value) triggering is selected. Whether
the peak value detection in automatic trigger mode is
automatically activated or not, depends on the trigger
coupling setting (TRIG.MODE). The way the trigger point
symbol in the readout responds on different LEVEL
control knob settings indicates the situation:

1. If the trigger symbol can not be shifted in the vertical

direction when a signal is not applied or the signal
height is not sufficient, the peak value detection is
active.

2. Under the condition that the trigger point symbol

cannot be shifted in such a way that it leaves the signal
display on the screen, the peak value detection is
active.

3. The peak value detection is switched off if the trigger

point can be set outside the maximum peak values of
the signal, thus causing an untriggered signal display.

 Slope selection:

Briefly pressing this pushbutton selects which slope of
the signal is used for triggering the timebase generator.

Summary of Contents for HM 1004-3 Series

Page 1: ...Instruments Oscilloscope HM 1004 3 01 02 03 ENGLISH MANUAL HANDBUCH MANUEL...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...res 23 Phase difference measurement in DUAL mode Yt 24 Phase difference measurement in DUAL mode 24 Measurement of amplitude modulation 24 Triggering and timebase 25 Automatic Peak value Triggering 25...

Page 4: ...ollowing hints and conditions regarding emission and immunity should be observed 1 Data cables For the connection between instruments resp their interfaces and external devices computer printer etc su...

Page 5: ...x ambient temperature 0 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Color techno brown Cabinet W 285 H 125 D 380 mm Subject to change without notice 08 00 This microprocessor...

Page 6: ...f an extension cord without a protective conductor The mains line plug must be inserted before connec tions are made to measuring circuits The grounded accessible metal parts case sockets jacks and th...

Page 7: ...helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain intervals Only in this way is it largel...

Page 8: ...it a true signal display The display of sinusoidal signals within the bandwidth limits causes no prob lems but an increasing error in measurement due to gain reduction must be taken into account when...

Page 9: ...caus ing damage to the input of the oscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case...

Page 10: ...asurement of the rise or fall time is not limited to the trace dimensions shown in the above diagram It is only particularly simple in this way In principle it is possible to measure in any display po...

Page 11: ...Hz There fore the derating curve of the attenuator probe type con cerned must be taken into account The selection of the ground point on the test object is important when displaying small signal volta...

Page 12: ...iver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile memories These can be used by the ope...

Page 13: ...with this control knob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a h...

Page 14: ...aid of the X POS control As the X expansion results in a higher timebase speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over Please note...

Page 15: ...hannel II followed by XY Except the cursor lines which may be active all other readout information including the trigger point symbol are switched off In addition to all trigger and timebase related c...

Page 16: ...l switched off NR high frequency noise rejected LF low pass filter cuts off frequencies above approx 1 5kHz TVL TV signal line pulse triggering trigger point symbol switched off TVF TV signal frame pu...

Page 17: ...e TIME DIV 24 control knob is operative only for this timebase The readout then only displays the A time coefficient The timebase settings for this condition are stored if the timebase mode is changed...

Page 18: ...switch and B vernier VAR LED on In the latter case the TIME DIV knob can be used in the same way as de scribed before under A timebase condition Underneath the front panel sector described above the...

Page 19: ...r related con trols assumes that the readout is visible and the component tester is switched off 35 MENU Pushbutton Pressing and holding the pushbutton activates the display of the MAIN MENU It contai...

Page 20: ...as one signal sum or difference As the result can only be determined if both calibrated deflection coefficients are equal the CHI II 36 selection function is deactivated In that case the readout indic...

Page 21: ...ted to the oscilloscope input s Check that the device under test is switched off and connect the test lead s to the test point s Then switch on the instrument and afterwards the device under test The...

Page 22: ...rm distortion Prerequisite for this HF compensation is a square wave generator with fast risetime typically 4ns and low output impedance approx 50 providing 0 2V at a frequency of approx 1MHz The cali...

Page 23: ...this operation mode is the DUAL XY 18 pushbutton please note section Con trols and Readout In XY mode the timebase is deactivated The signal applied to the input of channel II front panel marking HOR...

Page 24: ...ng for one period in div In the example illustrated t 3div and T 10div The phase difference in degrees is calculated from Relatively small phase angles at lower frequencies can be measured more accura...

Page 25: ...ode for the entry into difficult measuring problems e g when the test signal is unknown relating to amplitude frequency or shape Presetting of all parameters is now possible with automatic triggering...

Page 26: ...5div height The polarity of the synchronization pulse is critical for the slope selection If the displayed sync pulses are above the picture field contents leading edge positive going then the slope...

Page 27: ...vated As the external trigger signal applied at the TRIG EXT socket normally has no relation to the signal height of the displayed signal the trigger point symbol is switched off The external trigger...

Page 28: ...nt This allows the sweep to begin on any portion of a signal except the first millimeters If the display shows jitter it is possible to select triggering of the B timebase after the elapsed delay time...

Page 29: ...lugs and with test prod alligator clip or sprung hook are required The test leads are connected as described in section Con trols and Readout Test Procedure Caution Do not test any component in live c...

Page 30: ...r a transistor the figures b e and b c are important The figure e c can vary but a vertical line only shows short circuit condition These transistor test patterns are valid in most cases but there are...

Page 31: ...devices In case of disregard of the safety warnings contained in this manual HAMEG refuses any liability regarding personal injury and or damage of equip ment Operation The oscilloscope is supplied w...

Page 32: ...Subject to change without notice 32 Front Panel HM1004 3...

Page 33: ...33 Subject to change without notice...

Page 34: ...Subject to change without notice 34...

Page 35: ...MANUAL HANDBUCH MANUEL...

Page 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: