Hameg HM 1004-3 Series Manual Download Page 19

19

Subject to change without notice

Controls and Readout

Input coupling:

Briefly pressing this pushbutton switches over from AC
(

symbol) to DC (

symbol) input coupling and vice versa.

The setting is displayed in the readout with the deflection
coefficient.

Probe factor:

Pressing and holding the pushbutton selects the indi-
cated deflection coefficient of channel II displayed in the
readout, between 1:1 and 10:1. In condition 10:1 the
probe factor is thus indicated by a probe symbol displayed
by the readout in front the channel information (e.g.
“probe symbol”, Y2...). In the case of cursor voltage
measurement, the probe factor is automatically included.

Please note:
The symbol must not be activated unless a x10 (10:1)
attenuator probe is used.

(33)GD - INV. 

- Pushbutton with two functions.

GD:

Each time this pushbutton is pressed briefly, the input is
switched from active to inactive and vice versa. It is
displayed in the readout as an earth (ground) symbol
instead of the deflection coefficient and the

 ~ 

(AC) or

 =

(DC) symbol.

The GD setting disables the input signal, the

 AC/DC (32)

pushbutton and the

 VOLTS/DIV (20) 

knob. In automatic

trigger mode the undeflected trace is visible representing
the 0 Volt trace position.

The ”0 Volt” position is indicated by a ”

” symbol in the

readout. Please note 

Y-POS. II (10)

.

INV.:

Pressing and holding this pushbutton switches the chan-
nel II invert function on or off. The invert “on” condition
is indicated by the readout with a horizontal bar above

“Y2” 

(Yt mode). The invert function causes the signal

display of channel II to be inverted by 180°.

In XY mode the invert function of channel II is auto-
matically switched off.

(34)TRIG.  EXT.

This BNC socket is the external trigger signal input. Briefly
pressing the

 TRIG. (19) 

pushbutton, until the information

“TR:EXT, slope, coupling” 

is visible in the readout and

the TRIG. “EXT” -LED is lit, switches the input active. The
outer (ground) connection is galvanically connected to the
instrument ground and consequently to the safety earth
contact of the line/mains plug.

Below the CRT there are the controls for the readout,
the component tester and the squarewave calibrator
with their outputs.

Please note:
The following description of the cursor related con-
trols assumes that the readout is visible and the
component tester is switched off.

(35) MENU 

- Pushbutton

Pressing and holding the pushbutton activates the display
of the 

MAIN MENU

. It contains the submenus 

SETUP

and 

CALIBRATE

.

Once a menu is displayed, the following pushbuttons are
of importance:

1. 

SAVE

 and 

RECALL

 (7) pushbutton.

Briefly pressing selects the submenu or an item within
the submenu.

2. 

SAVE

 (7) pushbutton 

SET

 function.

Pressing and holding the 

SAVE

 (7) pushbutton calls (sets)

the menu or the previously selected item. In those cases
where the item is marked ON/Off the setting changes
from 

ON

 to 

OFF

 or vive versa.

If "

PRESS SECURITY KEY

" is displayed, this function is

available only for HAMEG authorized workshops. Press

AUTOSET

 (3) to switch back to the menu.

3. 

AUTOSET

 (3) pushbutton.

Each time the 

AUTOSET

 pushbutton is pressed the

menu switches back one step until 

MAIN MENU

 is

displayed. Then pressing the 

AUTOSET

 pushbutton again

switches the menu operation off and 

AUTOSET 

is set to

the normal function.

(36)ON/OFF - CHI/II  - 1/

∆∆∆∆∆

t

This pushbutton has several functions.

ON/OFF:

Pressing and holding the pushbutton switches both cur-
sor lines  on or off. As the cursor lines are part of the
readout, they are visible only if the readout is switched on.

CHI/II:

This function     

is required and available only in DUAL

and 

XY

     mode     

in combination with 

∆∆∆∆∆

V (38)

measurement. Briefly pressing the button switches
between channel I and channel II and vice versa.

In 

DUAL

 mode the measured result is displayed by the

readout with      “

∆∆∆∆∆

V1: ...

”     or      “

∆∆∆∆∆

V2: ...

 

providing the

deflection coefficient is calibrated. The settings of the
cursors must relate to the signal of the selected channel.

In 

XY

 mode the instrument is automatically set to 

∆∆∆∆∆

V

measurement. In this mode two signals are normally
applied causing an X and a Y deflection. The deflection
coefficient selected for each channel may be different,
thus as in     

DUAL

     mode the 

∆∆∆∆∆

V

 cursor measurement

requires a channel selection. Under channel I (Y signal)
measuring condition the cursor lines are displayed as
horizontal lines and the readout displays

 

∆∆∆∆∆

VY: ...

(calibrated). Briefly pressing the pushbutton changes to
channel II (X signal) voltage measurement. Then the cursor
lines are displayed as vertical lines and the readout
indicates     “

∆∆∆∆∆

VX> ..

.

”     (uncalibrated).

In

 CH I 

or

 CH II 

mono mode, only one deflection coeffi-

cient is present and there is no requirement to select

Summary of Contents for HM 1004-3 Series

Page 1: ...Instruments Oscilloscope HM 1004 3 01 02 03 ENGLISH MANUAL HANDBUCH MANUEL...

Page 2: ...MANUAL HANDBUCH MANUEL...

Page 3: ...res 23 Phase difference measurement in DUAL mode Yt 24 Phase difference measurement in DUAL mode 24 Measurement of amplitude modulation 24 Triggering and timebase 25 Automatic Peak value Triggering 25...

Page 4: ...ollowing hints and conditions regarding emission and immunity should be observed 1 Data cables For the connection between instruments resp their interfaces and external devices computer printer etc su...

Page 5: ...x ambient temperature 0 C 40 C Protective system Safety class I IEC1010 1 Weight approx 5 9kg Color techno brown Cabinet W 285 H 125 D 380 mm Subject to change without notice 08 00 This microprocessor...

Page 6: ...f an extension cord without a protective conductor The mains line plug must be inserted before connec tions are made to measuring circuits The grounded accessible metal parts case sockets jacks and th...

Page 7: ...helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain intervals Only in this way is it largel...

Page 8: ...it a true signal display The display of sinusoidal signals within the bandwidth limits causes no prob lems but an increasing error in measurement due to gain reduction must be taken into account when...

Page 9: ...caus ing damage to the input of the oscilloscope However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case...

Page 10: ...asurement of the rise or fall time is not limited to the trace dimensions shown in the above diagram It is only particularly simple in this way In principle it is possible to measure in any display po...

Page 11: ...Hz There fore the derating curve of the attenuator probe type con cerned must be taken into account The selection of the ground point on the test object is important when displaying small signal volta...

Page 12: ...iver please note trace rotation TR 6 FOCUS This control knob effects both the trace and the readout sharpness 7 SAVE RECALL The instrument contains 9 non volatile memories These can be used by the ope...

Page 13: ...with this control knob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to XY mode this control knob is inactive and the X POS knob must be used for a h...

Page 14: ...aid of the X POS control As the X expansion results in a higher timebase speed lower time deflection coefficient all time and frequency relevant information in the readout is switched over Please note...

Page 15: ...hannel II followed by XY Except the cursor lines which may be active all other readout information including the trigger point symbol are switched off In addition to all trigger and timebase related c...

Page 16: ...l switched off NR high frequency noise rejected LF low pass filter cuts off frequencies above approx 1 5kHz TVL TV signal line pulse triggering trigger point symbol switched off TVF TV signal frame pu...

Page 17: ...e TIME DIV 24 control knob is operative only for this timebase The readout then only displays the A time coefficient The timebase settings for this condition are stored if the timebase mode is changed...

Page 18: ...switch and B vernier VAR LED on In the latter case the TIME DIV knob can be used in the same way as de scribed before under A timebase condition Underneath the front panel sector described above the...

Page 19: ...r related con trols assumes that the readout is visible and the component tester is switched off 35 MENU Pushbutton Pressing and holding the pushbutton activates the display of the MAIN MENU It contai...

Page 20: ...as one signal sum or difference As the result can only be determined if both calibrated deflection coefficients are equal the CHI II 36 selection function is deactivated In that case the readout indic...

Page 21: ...ted to the oscilloscope input s Check that the device under test is switched off and connect the test lead s to the test point s Then switch on the instrument and afterwards the device under test The...

Page 22: ...rm distortion Prerequisite for this HF compensation is a square wave generator with fast risetime typically 4ns and low output impedance approx 50 providing 0 2V at a frequency of approx 1MHz The cali...

Page 23: ...this operation mode is the DUAL XY 18 pushbutton please note section Con trols and Readout In XY mode the timebase is deactivated The signal applied to the input of channel II front panel marking HOR...

Page 24: ...ng for one period in div In the example illustrated t 3div and T 10div The phase difference in degrees is calculated from Relatively small phase angles at lower frequencies can be measured more accura...

Page 25: ...ode for the entry into difficult measuring problems e g when the test signal is unknown relating to amplitude frequency or shape Presetting of all parameters is now possible with automatic triggering...

Page 26: ...5div height The polarity of the synchronization pulse is critical for the slope selection If the displayed sync pulses are above the picture field contents leading edge positive going then the slope...

Page 27: ...vated As the external trigger signal applied at the TRIG EXT socket normally has no relation to the signal height of the displayed signal the trigger point symbol is switched off The external trigger...

Page 28: ...nt This allows the sweep to begin on any portion of a signal except the first millimeters If the display shows jitter it is possible to select triggering of the B timebase after the elapsed delay time...

Page 29: ...lugs and with test prod alligator clip or sprung hook are required The test leads are connected as described in section Con trols and Readout Test Procedure Caution Do not test any component in live c...

Page 30: ...r a transistor the figures b e and b c are important The figure e c can vary but a vertical line only shows short circuit condition These transistor test patterns are valid in most cases but there are...

Page 31: ...devices In case of disregard of the safety warnings contained in this manual HAMEG refuses any liability regarding personal injury and or damage of equip ment Operation The oscilloscope is supplied w...

Page 32: ...Subject to change without notice 32 Front Panel HM1004 3...

Page 33: ...33 Subject to change without notice...

Page 34: ...Subject to change without notice 34...

Page 35: ...MANUAL HANDBUCH MANUEL...

Page 36: ...uk Spain HAMEG S L Villarroel 172 174 08036 BARCELONA Tel f 93 4301597 Telefax 93 321220 E mail email hameg es France HAMEG S a r l 5 9 av de la R publique 94800 VILLEJUIF T l 1 4677 8151 Telefax 1 4...

Reviews: