6
Warning
:
Do not rely on the SELECT-A-FLOW* dial key
or tie-wrap to prevent patient tampering.
SELECT-A-FLOW* device is available in two flow rate ranges
and distinquished by color on the face of device:
Blue
(1-7 ml/hr)
•
Flow rate: 1, 2, 3, 4, 5, 6, 7 ml/hr
Purple
(2-14 ml/hr)
•
Flow rate: 2, 4, 6, 8, 10, 12, 14 ml/hr
Instructions for Use
Use Aseptic Technique
Filling The ON-Q* Pump: (Figure 4)
Note:
Follow hospital protocols and applicable regulations for
filling pump.
1. Close clamp.
2. Uncap the fill port.
3. Attach filled syringe to fill port. Invert pump as shown.
4. Grasp syringe with both hands.
5. Push down on plunger continuously until volume is
dispensed. Do not handle pump while filling, as the
syringe tip may break. Repeat as necessary. Syringe
accuracy is ±4%.
Note:
Filling Extension Sets are provided with larger
pumps (see product insert).
6. Remove filling device from fill port
7. Replace fill port cap. Label with appropriate
pharmaceutical and patient information.
Figure 4
Fill Volume
Caution:
Do not underfill pump. Underfilling the pump
may significantly increase the flow rate. Do not exceed
maximum fill volume. (Table 1)
Table 1: Fill Volume (ml)
Fill Volume
100
200 270
400
600
Max Fill Volume
125
335 335
550
750
Retained Volume
≤5
≤11 ≤11 ≤16 ≤23
Warning: Flow rate is adjustable. Medication dosage
should be based on maximum flow rate. Pump is pre-set
at 7 ml/hr or 14 ml/hr (depending on model).
Table 2: 24-Hour Dosing Reference
SELECT-A-FLOW*
Model
Drug Concentration (mg)
0.2%
0.25%
0.5%
7 ml/hr Max Flow Rate
336
420
840
14 ml/hr Max Flow Rate
672
840
1680
Formula:
ml/hr x % drug concentration x 10 x 24 hr = 24 hour dose (mg)
Caution:
Calculations based on the labeled flow rate.
Flow rate accuracy varies.
Priming the Administration Set
Use Aseptic Technique
Note:
SELECT-A-FLOW* device is packaged with the flow rate
at the highest setting to minimize priming time.
1. Open the plastic cover of the SELECT-A-FLOW* device
(Figure 5)
2. To begin priming the SELECT-A-FLOW* device, ensure
highest flow rate setting is selected.
• Make sure the selected flow rate is aligned within the
window below the ml/hr
x
mark.
• The tactile feel will allow the user to ensure selected
flow rate is set.
3. Open the clamp and remove tubing cap to begin priming.
Summary of Contents for ON-Q Select-A-Flow
Page 1: ...PUMP WITH SELECT A FLOW VARIABLE RATE CONTROLLER Instructions for Use...
Page 2: ......
Page 4: ...2...
Page 46: ...44 PVC 2 DEHP DEHP DEHP DEHP PVC 1 Halyard Health...
Page 94: ...92 2 1 Halyard Health...
Page 98: ...96 3 5 6 7 4 5 6 6 1 2 SELECT A FLOW SELECT A FLOW 7 8 3 8 3 6 9 9...
Page 123: ...121 1 DEHP PVC DEHP DEHP DEHP PVC 1 Halyard Health SELECT A FLOW...
Page 130: ...128 DEHP 2 PVC DEHP DEHP DEHP PVC 1 Halyard Health...
Page 140: ...138 5 6 7 4 5 6 6 1 2 SELECT A FLOW SELECT A FLOW ml 7 8 3 8 3 ml 6 ml ml 9 9 10 10...
Page 142: ...140...
Page 143: ...141...