5.
For analyzers with the optional cationic pump, install the outlet tube from the cationic pump back
into the off-center hole in the cap.
6.
Close the analytics panel. Turn the latch on the analytics panel to the locked position.
7.
Enter the bottle level as follows:
a.
Push
menu
, then select REAGENTS/STANDARDS.
b.
Select SET COND LEVEL, then enter 100%.
8.
Push
menu
, then select START ANALYZER.
Replace the KCl electrolyte
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
Replace the KCl electrolyte in the KCl electrolyte reservoir at 3-month intervals.
Items to collect:
• Personal protective equipment (refer to MSDS/SDS)
• 3M KCl electrolyte, 200 mL
• Small container, 250 mL or larger
Replace the KCl electrolyte as follows:
1.
Put on the personal protective equipment identified in the safety data sheet (MSDS/SDS).
2.
Drain the contents of the KCl electrolyte reservoir into the small container. Refer to the illustrated
steps that follow. Discard the KCl electrolyte into an applicable drain.
8
English
Summary of Contents for NA5600 sc Na+
Page 7: ...English 7...
Page 13: ...Remove the reference electrode Refer to the illustrated steps that follow English 13...
Page 33: ...Deutsch 33...
Page 38: ...38 Deutsch...
Page 61: ...Italiano 61...
Page 66: ...66 Italiano...
Page 88: ...88 Fran ais...
Page 93: ...Fran ais 93...
Page 94: ...Retrait de l lectrode de r f rence Reportez vous aux tapes illustr es suivantes 94 Fran ais...
Page 115: ...Espa ol 115...
Page 120: ...120 Espa ol...
Page 142: ...142 Portugu s...
Page 147: ...Portugu s 147...
Page 148: ...Remover o el ctrodo de refer ncia Consulte os passos ilustrados abaixo 148 Portugu s...
Page 171: ...Nederlands 171...
Page 176: ...176 Nederlands...
Page 177: ...Referentie elektrode verwijderen Volg de volgende afgebeelde stappen Nederlands 177...
Page 198: ...198 Polski...
Page 203: ...Polski 203...
Page 204: ...Wyjmowanie elektrody odniesienia Sprawdzi list kolejnych czynno ci na rysunku 204 Polski...
Page 224: ...224 Svenska...
Page 230: ...Ta bort referenselektroden Se de illustrerade stegen som f ljer 230 Svenska...
Page 250: ...250 Suomi...
Page 256: ...Viite elektrodin irrottaminen Lue seuraavat kuvaohjeet 256 Suomi...
Page 277: ...Magyar 277...
Page 282: ...282 Magyar...
Page 283: ...A referenciaelektr da elt vol t sa L sd az al bbi k peken bemutatott l p seket Magyar 283...
Page 301: ...1 1 3 301 X 303 X KCl 306 X 307 X 1 2 MSDS SDS 3 10 301...
Page 303: ...1 DIPA DIPA 303...
Page 305: ...305...
Page 308: ...T 1 6 A 250 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 308...
Page 309: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 309...
Page 310: ...1 250 1 5 2x 2x KCl MSDS SDS 310...
Page 311: ...311...
Page 312: ...312...
Page 313: ...DIPA DIPA MSDS SDS 313...
Page 314: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 314...
Page 322: ...KCl2 1 4 KCl 322 SET FUNCTION FEEDER CONTROL diag OVERFEED RESET 2 KCl 1 322...
Page 323: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 323...
Page 335: ...T rk e 335...
Page 340: ...340 T rk e...
Page 341: ...Referans elektrotunun kar lmas A a da g sterilen resimli ad mlara bak n T rk e 341...
Page 359: ...1 1 3 359 X 361 X KCl 364 X 365 X 1 2 MSDS SDS 3 10 359...
Page 361: ...1 DIPA DIPA 361...
Page 363: ...363...
Page 366: ...T 1 6 A 250 VAC 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 366...
Page 367: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 367...
Page 368: ...1 250 mL 1 5 L 2x 2x KCl MSDS SDS 368...
Page 369: ...369...
Page 370: ...370...
Page 371: ...DIPA DIPA MSDS SDS 371...
Page 372: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 372...
Page 377: ...PH TOO LOW pH 10 361 o diag PERFORM TEST CONDITIONING VALVE 377...
Page 382: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 382...
Page 389: ......