![Hach NA5600 sc Na+ Maintenance And Troubleshooting Manual Download Page 23](http://html2.mh-extra.com/html/hach/na5600-sc-na/na5600-sc-na_maintenance-and-troubleshooting-manual_3331134023.webp)
1.
Push
diag
, then select PERFORM TEST.
2.
Select an option.
Option
Description
REACT SOL
DELIVERY
Starts the reactivation solution pump. The reactivation solution flows into the middle
chamber of the measurement cell.
Note: The measurement cell is the clear block where the electrodes are installed.
SAMPLE DELIVERY
Sets the sample solenoid valve for the selected channel to open. The sample flows
into the overflow vessel.
Note: If the overflow vessel is full, the level of the water in the overflow vessel will
not change. To drain the overflow vessel, do the SAMPLE DRAIN test.
CAL STD DELIVERY
Note: The CAL STD DELIVERY option is only available on analyzers that have the
auto calibration option.
Sets the calibration solution pump to on. The calibration solution flows into the
overflow vessel.
CONDITIONING
VALVE
Sets the conditioning valve to open. The conditioning gas flows into the right
chamber of the measurement cell.
Note: The measurement cell is the clear block where the electrodes are installed.
SAMPLE DRAIN
Sets the drain pump to on. The water in the overflow vessel drains out the chemical
drain tube.
MIXER
Sets the mixer motor to on. The stir bar in the overflow vessel turns.
STATUS LED
Sets the status indicator light to continuously cycle: yellow, red, green.
AIR PUMP
Note: The AIR PUMP option is only available when the optional cationic pump is
installed.
Sets the cationic pump to on. The cationic pump pushes air into the conditioning
solution (DIPA) bottle. Refer to the
Product overview
section of the installation
manual to identify the location of the cationic pump in the analyzer.
SELECT SCRIPT
Note: The SELECT SCRIPT option is for Service use only.
Sets the analyzer to use the normal instrument script or the test script.
Start a 4–20 mA output or relay test
Start a 4–20 mA output or relay test to identify if a 4–20 mA output or relay operation is correct.
1.
Push
diag
and select OUTPUTS.
2.
Select an option.
Option
Description
TEST 4–20 mA
Sets the signal (mA) at the selected 4–20 mA output. Measure the current at the
4–20 mA output with a multimeter.
TEST RELAY
Sets the selected relay to on (energized) or off (not energized). Measure the
resistance across the relay with a multimeter to identify if the relay is open or closed.
HOLD OUTPUTS
Holds the outputs at the last value or sets them to the transfer value.
ACTIVATION
—Select LAUNCH to hold the outputs at the last value or set them to the transfer
value. Select RELEASE to let the outputs operate normally.
SET OUTMODE
—
Select HOLD OUTPUTS to hold the outputs at the last value. Select TRANSFER
OUTPUTS to set the outputs to the transfer value. Refer to the operations manual to
set the transfer value for the outputs.
SET CHANNELS
—Select ALL to select all of
the channels. Select ANALYZER to select all of the channels for the selected
analyzer.
English
23
Summary of Contents for NA5600 sc Na+
Page 7: ...English 7...
Page 13: ...Remove the reference electrode Refer to the illustrated steps that follow English 13...
Page 33: ...Deutsch 33...
Page 38: ...38 Deutsch...
Page 61: ...Italiano 61...
Page 66: ...66 Italiano...
Page 88: ...88 Fran ais...
Page 93: ...Fran ais 93...
Page 94: ...Retrait de l lectrode de r f rence Reportez vous aux tapes illustr es suivantes 94 Fran ais...
Page 115: ...Espa ol 115...
Page 120: ...120 Espa ol...
Page 142: ...142 Portugu s...
Page 147: ...Portugu s 147...
Page 148: ...Remover o el ctrodo de refer ncia Consulte os passos ilustrados abaixo 148 Portugu s...
Page 171: ...Nederlands 171...
Page 176: ...176 Nederlands...
Page 177: ...Referentie elektrode verwijderen Volg de volgende afgebeelde stappen Nederlands 177...
Page 198: ...198 Polski...
Page 203: ...Polski 203...
Page 204: ...Wyjmowanie elektrody odniesienia Sprawdzi list kolejnych czynno ci na rysunku 204 Polski...
Page 224: ...224 Svenska...
Page 230: ...Ta bort referenselektroden Se de illustrerade stegen som f ljer 230 Svenska...
Page 250: ...250 Suomi...
Page 256: ...Viite elektrodin irrottaminen Lue seuraavat kuvaohjeet 256 Suomi...
Page 277: ...Magyar 277...
Page 282: ...282 Magyar...
Page 283: ...A referenciaelektr da elt vol t sa L sd az al bbi k peken bemutatott l p seket Magyar 283...
Page 301: ...1 1 3 301 X 303 X KCl 306 X 307 X 1 2 MSDS SDS 3 10 301...
Page 303: ...1 DIPA DIPA 303...
Page 305: ...305...
Page 308: ...T 1 6 A 250 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 308...
Page 309: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 309...
Page 310: ...1 250 1 5 2x 2x KCl MSDS SDS 310...
Page 311: ...311...
Page 312: ...312...
Page 313: ...DIPA DIPA MSDS SDS 313...
Page 314: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 314...
Page 322: ...KCl2 1 4 KCl 322 SET FUNCTION FEEDER CONTROL diag OVERFEED RESET 2 KCl 1 322...
Page 323: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 323...
Page 335: ...T rk e 335...
Page 340: ...340 T rk e...
Page 341: ...Referans elektrotunun kar lmas A a da g sterilen resimli ad mlara bak n T rk e 341...
Page 359: ...1 1 3 359 X 361 X KCl 364 X 365 X 1 2 MSDS SDS 3 10 359...
Page 361: ...1 DIPA DIPA 361...
Page 363: ...363...
Page 366: ...T 1 6 A 250 VAC 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 366...
Page 367: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 367...
Page 368: ...1 250 mL 1 5 L 2x 2x KCl MSDS SDS 368...
Page 369: ...369...
Page 370: ...370...
Page 371: ...DIPA DIPA MSDS SDS 371...
Page 372: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 372...
Page 377: ...PH TOO LOW pH 10 361 o diag PERFORM TEST CONDITIONING VALVE 377...
Page 382: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 382...
Page 389: ......