Problem
Possible cause
Solution
The instrument
readings are not stable.
The analyzer was started
less than 2 hours ago.
Let the analyzer operate for 2 hours to become stable.
The electrode cable(s) are
cross-threaded on the
electrodes.
Disconnect the cable from the electrode. Push the
connector fully down on the electrode connector, then turn
the connector until finger tight.
The electrode cables are
connected to the wrong
electrodes.
Connect the blue cable to the reference electrode.
Connect the black cable to the sodium electrode.
The KCl electrolyte flow
rate is too fast.
Remove the reference electrode from the measurement
cell.
Push the collar down and turn the collar less than 1/4 of a
turn to lock the collar. When the collar is locked, the collar
does not turn. If the collar is not locked, the KCl electrolyte
will flow too fast from the reference electrode into the
measurement cell.
The measurements are
not accurate.
The analyzer is not
calibrated.
Calibrate the analyzer. Refer to
Do a calibration
in the
operations manual. For the best accuracy, calibrate the
analyzer at 7-day intervals (weekly).
Auto reactivation is
disabled or the interval
between reactivations is
more than 24 hours.
Start an auto reactivation. Push
diag
, then select
SENSOR REACTIVATION.
Note: For the best accuracy, set auto reactivation to on. Set the
interval to 24 hours. Refer to the operations manual.
The reactivation solution
bottle is empty.
Replace the reactivation solution. Refer to
calibration and reactivation solutions
on page 4.
There is a leak or
blockage in the
reactivation solution
tubing.
Look for a blockage or leak in the tubing. Replace the
tubing as necessary.
The reactivation solution
pump does not operate
correctly.
Push
diag
, then select PERFORM TEST > REACT SOL
DELIVERY > HIGH to set the reactivation solution pump
to a high flow rate. If reactivation solution does not flow to
the middle chamber of the measurement cell, replace the
reactivation solution pump.
The stir bar in the
overflow vessel does
not turn.
The mixer does not
operate correctly.
Note: The stir bar turns
intermittently.
Push
diag
, then select PERFORM TEST > MIXER to set
the mixer motor to on. If the stir bar does not move,
replace the mixer motor.
The KCl electrolyte
consumption is too
fast
1
.
The collar on the glass
part of the reference
electrode is not locked.
Remove the reference electrode.
Push the collar up. Push the collar down and turn the
collar less than 1/4 of a turn to lock the collar. Refer to the
illustrated steps in
Install the reference electrode
in the
installation manual.
When the collar is locked, the collar does not turn. If the
collar is not locked, the KCl electrolyte will flow too fast
from the reference electrode into the measurement cell.
1
The KCl electrolyte consumption is too fast if the level of the reservoir decreases by more than
3 mm (1/8 in.) in two days
20
English
Summary of Contents for NA5600 sc Na+
Page 7: ...English 7...
Page 13: ...Remove the reference electrode Refer to the illustrated steps that follow English 13...
Page 33: ...Deutsch 33...
Page 38: ...38 Deutsch...
Page 61: ...Italiano 61...
Page 66: ...66 Italiano...
Page 88: ...88 Fran ais...
Page 93: ...Fran ais 93...
Page 94: ...Retrait de l lectrode de r f rence Reportez vous aux tapes illustr es suivantes 94 Fran ais...
Page 115: ...Espa ol 115...
Page 120: ...120 Espa ol...
Page 142: ...142 Portugu s...
Page 147: ...Portugu s 147...
Page 148: ...Remover o el ctrodo de refer ncia Consulte os passos ilustrados abaixo 148 Portugu s...
Page 171: ...Nederlands 171...
Page 176: ...176 Nederlands...
Page 177: ...Referentie elektrode verwijderen Volg de volgende afgebeelde stappen Nederlands 177...
Page 198: ...198 Polski...
Page 203: ...Polski 203...
Page 204: ...Wyjmowanie elektrody odniesienia Sprawdzi list kolejnych czynno ci na rysunku 204 Polski...
Page 224: ...224 Svenska...
Page 230: ...Ta bort referenselektroden Se de illustrerade stegen som f ljer 230 Svenska...
Page 250: ...250 Suomi...
Page 256: ...Viite elektrodin irrottaminen Lue seuraavat kuvaohjeet 256 Suomi...
Page 277: ...Magyar 277...
Page 282: ...282 Magyar...
Page 283: ...A referenciaelektr da elt vol t sa L sd az al bbi k peken bemutatott l p seket Magyar 283...
Page 301: ...1 1 3 301 X 303 X KCl 306 X 307 X 1 2 MSDS SDS 3 10 301...
Page 303: ...1 DIPA DIPA 303...
Page 305: ...305...
Page 308: ...T 1 6 A 250 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 308...
Page 309: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 309...
Page 310: ...1 250 1 5 2x 2x KCl MSDS SDS 310...
Page 311: ...311...
Page 312: ...312...
Page 313: ...DIPA DIPA MSDS SDS 313...
Page 314: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 314...
Page 322: ...KCl2 1 4 KCl 322 SET FUNCTION FEEDER CONTROL diag OVERFEED RESET 2 KCl 1 322...
Page 323: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 323...
Page 335: ...T rk e 335...
Page 340: ...340 T rk e...
Page 341: ...Referans elektrotunun kar lmas A a da g sterilen resimli ad mlara bak n T rk e 341...
Page 359: ...1 1 3 359 X 361 X KCl 364 X 365 X 1 2 MSDS SDS 3 10 359...
Page 361: ...1 DIPA DIPA 361...
Page 363: ...363...
Page 366: ...T 1 6 A 250 VAC 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 366...
Page 367: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 367...
Page 368: ...1 250 mL 1 5 L 2x 2x KCl MSDS SDS 368...
Page 369: ...369...
Page 370: ...370...
Page 371: ...DIPA DIPA MSDS SDS 371...
Page 372: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 372...
Page 377: ...PH TOO LOW pH 10 361 o diag PERFORM TEST CONDITIONING VALVE 377...
Page 382: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 382...
Page 389: ......