94
SECTION 11,
continued
PRECAUÇÃO
Quando a Lâmpada estiver acesa e a proteção de lâmpada estiver removida, use protetor
ocular.
2. Remove the access panel to expose the lamp compartment (see Figure 22).
Squeeze the two latch tabs together, and pull out on the panel.
Figure 22
Lamp Replacement
3. Use a small screwdriver to loosen the two screws securing the lamp leads in
the lamp terminal block. Pull the leads free.
4. Remove the two Phillips screws securing the lamp holder in the instrument
(upper left-hand corner and lower right-hand corner). Remove the entire lamp
assembly.
Note: Do not touch the lamp; oil from skin will cause damage. Clean the lamp with alcohol
if necessary.
5. Secure the new lamp assembly in place using the two Phillips screws (do not
over-tighten).
6. Insert one stripped lamp lead under each of the positions in the lamp terminal
block. Either lamp lead can be inserted in either terminal block position.
Tighten the terminal block screws to secure the leads (do not over-tighten).
Replace the lamp compartment access panel and restore power to the
instrument.
Note: The lamp output is stabilized at Hach; no “burn-in” time is required. However,
warm-up the lamp for a minimum of 60 minutes before calibrating the instrument.
7. Recalibrate the instrument as described in Section 3.2.4 on page 39. New
turbidity standards must be prepared (refer to Section 3.2.2 on page 37 and
Section 3.2.3 on page 37).
Summary of Contents for Lange 2100AN
Page 6: ...6...
Page 8: ...8...
Page 10: ...10...
Page 14: ...14...
Page 46: ...46...
Page 48: ...48...
Page 64: ...64...
Page 74: ...74...
Page 76: ...76...
Page 78: ...78...
Page 84: ...84...
Page 92: ...92...
Page 102: ...102...
Page 103: ...103 HOW TO ORDER...
Page 106: ...106...
Page 107: ...107...