38
SECTION 3,
continued
Note: Install each cell with the orientation mark aligned with the cell holder reference mark
when using matched vials.
Table 2 Formazin Standard Preparation
Standard
Step 1
Step 2
Step 3
20 NTU
Add 100 mL of dilution water to a
clean 200-mL class A volumetric
flask
With a TenSette
®
* Pipet, add 1.00
mL of well-mixed 4000-NTU
Formazin stock solution to the
200-mL flask.
Dilute to the mark with dilution water.
Stopper and mix.
200 NTU
Add 50 mL of dilution water to a
clean 100-mL class A volumetric
flask
With a TenSette* Pipet, add 5.00 mL
of well-mixed 4000-NTU Formazin
stock solution to the 100-mL flask.
Dilute to the mark with dilution water.
Stopper and mix.
1000 NTU
Add 50 mL of dilution water to a
clean 100-mL class A volumetric
flask
With a TenSette* Pipet, add 25.00
mL of well-mixed 4000-NTU
Formazin stock solution to the
100-mL flask.
Dilute to the mark with dilution water.
Stopper and mix.
4000 NTU
Transfer approximately 30 mL of well
mixed 4000-NTU Formazin stock
solution to a clean sample cell. No
dilution is required.
7500 NTU
Ampule
Use in the sealed ampule as
provided. DO NOT open the vial or
use its contents as dilution stock.
* Class A volumetric pipets can be used in place of a TenSette Pipet.
Summary of Contents for Lange 2100AN
Page 6: ...6...
Page 8: ...8...
Page 10: ...10...
Page 14: ...14...
Page 46: ...46...
Page 48: ...48...
Page 64: ...64...
Page 74: ...74...
Page 76: ...76...
Page 78: ...78...
Page 84: ...84...
Page 92: ...92...
Page 102: ...102...
Page 103: ...103 HOW TO ORDER...
Page 106: ...106...
Page 107: ...107...