Connect the colorimeter cable as follows:
a.
Align and push the colorimeter connector on the mating connector.
b.
Lift the cable clamp.
c.
Push the connector into the cable clamp.
8.
Put the reagent delivery connector into the reagent delivery board. Refer to the next figure for the
location of the reagent delivery connector.
9.
Connect the power cord to the electrical outlet. Wait for the analyzer to start up (approximately
2 minutes).
10.
Complete the steps that show on the display to complete periodic maintenance. The display
screens are shown for reference in Appendix B.
English
51
Summary of Contents for CM130
Page 1: ...DOC023 53 80540 CM130 11 2018 Edition 5 User Manual...
Page 2: ......
Page 18: ...Figure 9 Analyzer dimensions 16 English...
Page 65: ...Figure 22 Remove the pump clamp English 63...
Page 77: ...English 75...
Page 79: ...English 77...
Page 80: ...78 English...
Page 81: ...English 79...
Page 82: ...80 English...
Page 83: ...English 81...
Page 84: ...82 English...
Page 85: ...English 83...
Page 86: ...84 English...
Page 88: ...86 English...
Page 90: ...88 English...
Page 92: ...90 English...
Page 94: ...92 English...
Page 95: ...English 93...
Page 96: ...94 English...
Page 98: ...96 English...
Page 99: ...English 97...
Page 100: ...98 English...
Page 101: ...English 99...
Page 102: ...100 English...
Page 103: ...English 101...
Page 104: ...102 English...
Page 105: ...English 103...
Page 106: ...104 English...
Page 107: ...English 105...
Page 111: ......