3.
Push the black tubing of the grab sample assembly into the sample inlet fitting. The sample inlet
fitting is behind the metal colorimeter inlet fitting.
4.
Plumb the sample inlet tubing as follows:
a.
Plumb one end of the coil of black tubing to the sample source between the primary and
secondary carbon tanks, as identified in ANSI/AAMI/ISO 13959
Water for hemodialysis and
related therapies
.
b.
Hold the open end of the black tubing over a drain, then open the valve at the sample source.
Rinse the tubing for 30 seconds.
c.
Push the open end of the coil of black tubing into the check valve of the grab sample
assembly.
Decrease the length of the black tubing to remove the unnecessary length. Leave at least 5 feet
of length.
Note: AAMI is the acronym for Association for the Advancement of Medical Instrumentation.
22
English
Summary of Contents for CM130
Page 1: ...DOC023 53 80540 CM130 11 2018 Edition 5 User Manual...
Page 2: ......
Page 18: ...Figure 9 Analyzer dimensions 16 English...
Page 65: ...Figure 22 Remove the pump clamp English 63...
Page 77: ...English 75...
Page 79: ...English 77...
Page 80: ...78 English...
Page 81: ...English 79...
Page 82: ...80 English...
Page 83: ...English 81...
Page 84: ...82 English...
Page 85: ...English 83...
Page 86: ...84 English...
Page 88: ...86 English...
Page 90: ...88 English...
Page 92: ...90 English...
Page 94: ...92 English...
Page 95: ...English 93...
Page 96: ...94 English...
Page 98: ...96 English...
Page 99: ...English 97...
Page 100: ...98 English...
Page 101: ...English 99...
Page 102: ...100 English...
Page 103: ...English 101...
Page 104: ...102 English...
Page 105: ...English 103...
Page 106: ...104 English...
Page 107: ...English 105...
Page 111: ......