Specifications are subject to change without notice.
Table 1 General specifications
Specification
Details
Dimensions
(W x H x D)
Analyzer: 37.0 x 53.8 x 22.6 cm (14.6 x 21.2 x 8.9 in.)
Remote indicator: 12.3 x 29.4 x 10.7 cm (4.8 x 11.6 x 4.2 in.)
Enclosure
Enclosure rating: IP52 rated with the door closed
Enclosure material: PC/ABS case, PC door, PC hinges and latches, 316 SST
hardware
Remote indicator material: PC/ABS
Display
10.9 cm (4.3 in.) color display, back-lit LCD, WQVGA
Weight
Analyzer: 8.9 kg (19.6 lbs)
Remote indicator: 1.1 kg (2.5 lbs)
Mounting
Analyzer: Wall
Remote indicator: Wall or ceiling
Protection class
I
Pollution degree
2
Installation category
II
Power requirements
Analyzer: 100 to 240 VAC, 50/60 Hz; 1 A maximum
Remote indicator: Power is supplied by the analyzer.
Power-loss alarm: Power is supplied by a lithium coin battery CR2032 (3 V) that is in
the remote indicator.
Recommended operating
temperature
15 to 30 °C (60 to 85 °F)
Safe operating
temperature
5 to 40 °C (41 to 104 °F)
Operating humidity
5 to 90% non-condensing at 30 °C (85 °F) maximum
Storage temperature
-20 to 60 °C (-4 to 140 °F)
Fittings
Sample line: ¼-in. OD push-to-connect fitting for plastic tubing
Drain line: slip-on fitting for ½-in. ID soft plastic tubing
Cabling
1
Ethernet: 100 m (328 ft) maximum, Ethernet cable with an RJ45 connector
Remote indicator: 762 m (2500 ft) maximum, CAT5e/CAT6 type Ethernet cable with
four twisted pairs of 24 AWG solid copper wiring.
Make sure that the cable shielding includes an inner foil wrap around all of the pairs
and an outer braid, also known as SFTP or SF/UTP. Do not use stranded or copper-
clad aluminum conductors.
Sample
Pressure: 276 to 689 kPa (40 to 100 psi) nominal
Pressure spikes: 827 kPa (120 psi) or less
Flow rate: 250 mL/minute minimum
Temperature: 15 to 30 °C (60 to 85 °F)
Filtration: 15 µm or less
Refer to
on page 72 interfering substances.
1
Cabling is supplied by the user.
English
3
Summary of Contents for CM130
Page 1: ...DOC023 53 80540 CM130 11 2018 Edition 5 User Manual...
Page 2: ......
Page 18: ...Figure 9 Analyzer dimensions 16 English...
Page 65: ...Figure 22 Remove the pump clamp English 63...
Page 77: ...English 75...
Page 79: ...English 77...
Page 80: ...78 English...
Page 81: ...English 79...
Page 82: ...80 English...
Page 83: ...English 81...
Page 84: ...82 English...
Page 85: ...English 83...
Page 86: ...84 English...
Page 88: ...86 English...
Page 90: ...88 English...
Page 92: ...90 English...
Page 94: ...92 English...
Page 95: ...English 93...
Page 96: ...94 English...
Page 98: ...96 English...
Page 99: ...English 97...
Page 100: ...98 English...
Page 101: ...English 99...
Page 102: ...100 English...
Page 103: ...English 101...
Page 104: ...102 English...
Page 105: ...English 103...
Page 106: ...104 English...
Page 107: ...English 105...
Page 111: ......