Remote indicator lights and sounds
The analyzer status and the chlorine status are identified with sounds and colored lights on the
remote indicator in the treatment room. Refer to
and
. Similar alert and alarm sounds
occur at the remote indicator and the analyzer.
When the battery life of the remote indicator is low, the remote indicator makes a continuous sound.
Replace the battery when the continuous sound is heard. Refer to
on page 60.
Note: Alarm and alert settings cannot be changed. Sound volumes and tones cannot be changed.
Table 3 Remote indicator—Light color descriptions
1
Chlorine status light
Off:
Chlorine is not being measured
Blue:
Chlorine is less than 0.10 mg/L as Cl
2
Red:
Chlorine is 0.10 mg/L or more as Cl
2
2
Analyzer status light
Off:
No communication with analyzer
Blue:
Normal operation
Amber:
Non-urgent problem with analyzer
Red:
Urgent problem with analyzer
3
Audio
Table 4 Remote indicator—Light combinations and sounds
Remote indicator
Status
Remote indicator
Status
Idle Mode
Chlorine:
Not being measured
Analyzer:
Normal operation
(slow flashing)
Audio:
No sound
Measurement Mode
Chlorine:
High chlorine
(≥ 0.10 mg/L)
4
Analyzer:
Normal operation
Audio:
Chlorine alarm tone
Measurement Mode
Chlorine:
Low chlorine
(< 0.10 mg/L)
4
Analyzer:
Normal operation
Audio:
No sound
Measurement Mode
Chlorine:
High chlorine
(≥ 0.10 mg/L)
4
Analyzer:
Analyzer alert
5
Audio:
Chlorine alarm tone
Measurement Mode
Chlorine:
Low chlorine
(< 0.10 mg/L)
4
Analyzer:
Analyzer alert
5
Audio:
No sound
Idle Mode
Chlorine:
Not being measured
Analyzer:
Analyzer alarm
6
Audio:
Analyzer alarm tone
Disconnected Mode
7
Chlorine:
Unknown
Analyzer:
Unknown
Audio:
Power-loss beep
8
4
Chlorine measurements continue
5
There is a non-urgent problem with the analyzer.
6
There is an urgent problem with the analyzer. The analyzer cannot make reliable
measurements.
7
There is no communication between the analyzer and the remote indicator. The analyzer does
not have power or the cable between the analyzer and the remote indicator is disconnected.
8
Stops after approximately 5 minutes. The power-loss beep is similar to the power-loss beep of a
smoke detector.
English
11
Summary of Contents for CM130
Page 1: ...DOC023 53 80540 CM130 11 2018 Edition 5 User Manual...
Page 2: ......
Page 18: ...Figure 9 Analyzer dimensions 16 English...
Page 65: ...Figure 22 Remove the pump clamp English 63...
Page 77: ...English 75...
Page 79: ...English 77...
Page 80: ...78 English...
Page 81: ...English 79...
Page 82: ...80 English...
Page 83: ...English 81...
Page 84: ...82 English...
Page 85: ...English 83...
Page 86: ...84 English...
Page 88: ...86 English...
Page 90: ...88 English...
Page 92: ...90 English...
Page 94: ...92 English...
Page 95: ...English 93...
Page 96: ...94 English...
Page 98: ...96 English...
Page 99: ...English 97...
Page 100: ...98 English...
Page 101: ...English 99...
Page 102: ...100 English...
Page 103: ...English 101...
Page 104: ...102 English...
Page 105: ...English 103...
Page 106: ...104 English...
Page 107: ...English 105...
Page 111: ......