Connect the remote indicator cable to the analyzer
Install the female blade terminal on the end of the remote indicator cable. Then, connect the remote
indicator cable to the remote indicator connector.
1.
Use the CAT5 cable stripper to remove 38 mm (1.5 in.) of the outer cable shielding from the
remote indicator cable.
2.
Untwist each pair of wires. Use the wire strippers to remove 7 mm (0.25 in.) of plastic from each
pair of twisted wires.
3.
Install the female blade terminal on the end of the remote indicator cable as follows:
a.
Twist the silver stranded (shield) wires together.
b.
Put the silver stranded wires into the female blade terminal.
c.
Use a crimper tool to close the female blade terminal on the silver stranded wires.
d.
Gently pull the silver stranded wires to make sure that the wires are held securely in the
female blade terminal.
If step
is not completed correctly, cut off the loose wires and do steps
40
English
Summary of Contents for CM130
Page 1: ...DOC023 53 80540 CM130 11 2018 Edition 5 User Manual...
Page 2: ......
Page 18: ...Figure 9 Analyzer dimensions 16 English...
Page 65: ...Figure 22 Remove the pump clamp English 63...
Page 77: ...English 75...
Page 79: ...English 77...
Page 80: ...78 English...
Page 81: ...English 79...
Page 82: ...80 English...
Page 83: ...English 81...
Page 84: ...82 English...
Page 85: ...English 83...
Page 86: ...84 English...
Page 88: ...86 English...
Page 90: ...88 English...
Page 92: ...90 English...
Page 94: ...92 English...
Page 95: ...English 93...
Page 96: ...94 English...
Page 98: ...96 English...
Page 99: ...English 97...
Page 100: ...98 English...
Page 101: ...English 99...
Page 102: ...100 English...
Page 103: ...English 101...
Page 104: ...102 English...
Page 105: ...English 103...
Page 106: ...104 English...
Page 107: ...English 105...
Page 111: ......