Oxygen analyser (ppb) - 9182 -
8-1
Chapter 8: Error messages
Λ
In case of errors, measurements are replaced by dashes "
- - -".
Error messages
Description
Error messages during a measurement
CONCENTRATION
TOO HIGH
The concentration value is out of the limits. Check the
current value and the calibration parameters.
STAND-BY
1.99 ppb
Electrode protection
In case of measuring range exceeding "concentration
too hugh" over 2 minutes, the error message "STAND-
BY" appears. From this moment on, the electrode
polarization voltage is switched off during 10 minutes.
After these 10 minutes, the voltage is switched on.
In STAND-BY the user may switch on the voltage at
any time by pressing ENTER.
CONCENTRATION
TOO LOW
The concentration value is out of the limits. Check the
current value and the calibration parameters.
TEMPERATURE
ERROR
The sample temperature is out of the limits. Check the
cable polarity. Check if there is any short-circuit or open
circuit.