Oxygen analyser (ppb) - 9182 -
1-4
ANALYSIS
Measuring range
0...2 ppm
Sensibility
≤
0,5 ppb
Reproducibility
± 0,5 ppb or ± 2% whichever the greater
Detection limit
≤
1 ppb
Response time (90 %)
≤
3 min
Ambient temperature
-20...60 °C (-4...140 °F)
Relative humidity
10...90%
TRANSMITTER
Display
Display in concentration units
Display of the direct, concentration
dependent cell current in µA
Display of the sample temperature in °C/°F
Programming via menus
Units
nA, µA, ppb-µg/l, ppm-mg/l, °C, °F
Calibration
Electrical zero, chemical zero, slope calibration
in the air, slope process calibration by
comparison with a laboratory measurement
Interferences
Any gas liable to reduce to a potential
> - 850 mV
Analogue outputs
2 x 0/4... 20 mA isolated from input signal, 800
Ohms load maximum
- for the measure or for the temperature
- mode: linear, bi-linear
- accuracy: 0,1 mA
Alarms
Number: 4
- Functions: alarm - system alarm - timer
- Hysteresis: 0...10%
- Delay: 0...999s
- Breaking power:
250 VAC, 3A maximum
30 VDC, 0,5A maximum
Use a cable (rated 105 °C and AWG22 to 14).
The external cable insulation should be cut as
close as possible from the terminal block.
Summary of Contents for POLYMETRON 9182
Page 1: ...Operator Manual POLYMETRON 9182 Oxygen analyzer ppb 221 191 082 Revision G 20 11 2007 ...
Page 2: ......
Page 8: ......
Page 39: ...Oxygen analyser ppb 9182 5 9 ...
Page 48: ......
Page 56: ......
Page 70: ......
Page 74: ......
Page 76: ......
Page 80: ......
Page 86: ......
Page 90: ...Oxygen analyser ppb 9182 A7 4 ...
Page 91: ......
Page 92: ......