Oxygen analyser (ppb) - 9182 -
5-16
5.1.4.6 mA ADJUST Menu
The analogue output signals are factory-adjusted (upper limit: 20 mA). However if you drift of the
20 mA appears on one of the outputs, the mA ADJUST menu enables you to correct the span
value. Connect a amperometer in serial to the analogue output terminals and adjust the value till
you read 20.0 mA on the amperemeter.
+
The value displayed does not correspond to a mA value.
Enter
OUTPUT 2
ADJUST mA
Select
OUTPUT 1
VALUE
ADJUST mA
-
0000
+