Oxygen analyser (ppb) - 9182 -
7-2
Proceed as follows for the assembling:
A.
Screw the membrane
†
on the body till the dead stop
ƒ
.
B.
Fill in the probe body with 5 ml of electrolyte by means of the syringe.
ð
Check there is no impurity or bubble in the electrolyte
.
C.
Insert the electrode
‚
in the probe body until you meet a resistance. Do not force.
D.
Complete the mounting by tightening the nut.
E.
Mount the filling screw and its gasket. Tight it sufficiently to avoid any leaking of electrolyte or
external pollution from sample.